Figure 36: Fit the filter box on the measuring device
1
Gas filter box
2
Reflector enclosure
3
Adapter plate for SR-unit
4
Sender/receiver unit
5
Intermediate enclosure
6
Pin
Carry out filter box measurement
1
Switch filter box on.
NOTE
The warming-up phase takes about 2.5 hours.
2
Make the settings on the filter box:
w
Set the sample gas chamber to “Zero gas”.
w
Set the “Measuring/Purging” valve to “Purging”.
3
Make the settings on the EvU:
w
Switch to Calibration mode with button “cal”.
w
Select menu “Box measuring” -> “gas”.
w
Enter password “1 2 3 4”.
The measuring device starts a zero adjust and then switches to operating mode
“Box measuring”.
4
Connect the test gas on the filter box.
w
Set the primary pressure to approx. 1000 hPa (1 bar)
5
Measure the individual chambers and/or chamber combinations of the filter box
w
Set the valves for the respective filter chambers to “Test gas”
w
Set the valve measure/purge to “Purging” for 2 to 3 minutes (until the mea‐
sured value has stabilized to a level).
w
Then set the valve to “Measuring”.
The overpressure from the purge phase now dissipates.
w
When the measured value has stabilized again, read off and note the value.
6
Leave operating mode “Box measuring” on the EvU with the return button.