Menu level
Explanation
Reset System
Restarts the measuring system
Reset Parameter
Resets parameters to factory settings
a
All settings are overwritten!
8.4
Operating using the serial interface
NOTE
Menu texts are shown in
English
.
NOTICE
Operating is only possible via the serial interface during operation in Ex area. Local
operating via the evaluation unit is possible when the Ex protection is not effective.
8.4.1
Significance of command line elements
Table 14: Significance of command line elements
Element
Meaning
Text style used
SAMPLE
Shows the name of the command or utility pro‐
gram.
CAPITAL LETTERS
{variable}
Shows several options from which the user
must select one, more or all.
Small letters in {round brack‐
ets}
[option]
Shows optional elements.
Small letters in [square brack‐
ets]
. , : ;
Punctuation marks are part of the command
and to be inserted as such.
Small letters
<cr>
Stands for pressing Enter (on the computer
keyboard).
Small letters
8.4.2
Setting for serial interfaces
Table 15: Standard setting for serial interface
Parameter
Description / value
Baudrate
9600 Baud
Data bits
7
Parity
Even parity
Stop bits
1
Protocol
No protocol
Interface line
1:1
Connection
2-2 / 3-3 / 5-5
8.4.3
Serial interface command - MEAS
Start measuring operation/switch to measuring operation
Syntax:
GM700<<GOTO MEAS
8
OPERATION
64