Menu level
Explanation
Analog In
Input 1 T
Unit: °C, K, °F
Live Zero: 0, 2, 4 mA
Range low: 0 ... 15000
Range high: 0 ... 15000
Input 2 p
Unit: hPa
Range low: 0 ... 15000
Range high: 0 ... 15000
Regress Funct.
Span
0.50 ... 1.99
Zero
±999
Ambient Temp.
Range
•
0 ... 50 °C
•
–10 ... 40 °C
•
–20 ... 30 °C
•
–30 ... 20 °C
[1]
•
–40 ... 10 °C
[1]
NOTICE
[1]
These temperature ranges are not allowed
for Ex applications.
t (Feed Test Gas)
Period
0 ... 3666 d (t = days), only when permanent cell
is included
Check cycle
Period
0 ... 24 h
Zero Adjust
Enable
Yes, No
Repet.
0 ... 24 h (only GPP)
t (purge)
0 ... 900 s; purge time until the process gas con‐
centration is zero
t (delay)
0 ... 1800 s; delay until gas enters again
Delta T
0 ... 999 °C; temperature difference at which a
zero adjust is carried out
Device Serial Number
Head
_ _ _ _ _ _ _ _
Laser
_ _ _ _ _ _ _ _
EvU
_ _ _ _ _ _ _ _
Software Revision
■
GMM700-X
■
XXXXXXX XXXX
■
GMM700/DSP
■
XXXXXXX XXXX
Service
Not defined
8
OPERATION
62
O P E R A T I N G I N S T R U C T I O N S | GM700 Ex
8019532/100Q/V1-1/2018-08 | SICK
Subject to change without notice