WARNING
Risk of injury when the device drops down
The weight of the device can cause it to drop down and cause injuries during the work
described in this chapter.
b
Carry out assembly work on parts of the device together with another person when
necessary.
Ex-suitable measuring probe
DANGER
Risk of explosion with a measuring probe not suitable for Ex zones
Measuring probes not approved for operation in potentially explosive atmospheres can
cause an explosion.
b
Only use the GMP measuring probe included in the scope of delivery from SICK.
5.2
Tools required
Tools
Part No.
Required for
Adjustment device
2034121
Alignment of "flanges with tube"
19 mm jaw wrench
Flange screw fitting
Screwdriver
Connections
Allen key
Connections
5.3
Material required
Material
Required for
Personal protective equipment
Protection when working at the sampling point
Tubes for protective gas
Protective gas feed line,
see "Connecting the protective gas
5.4
Preparing the sampling point
The operating company is responsible for preparing the sampling point
NOTICE
Basis for determining the sampling point:
•
Preceding project planning (e.g., based on the SICK application questionnaire)
•
Regulations of the local authorities
Responsibility of the operator
•
Determine sampling point
•
Preparing the sampling point
•
Feed and drain for protective gas
MOUNTING
5