Specifications
FLOWSIC600-XT · Operating Instructions · 8018846/1BHP/V2-1/2021-05 · © SICK Engineering GmbH
123
Su
bje
ct to
ch
an
ge
w
ith
ou
t n
otic
e
Table 18
Volume conversion
Installation
Dimensions (W x H x D)
See dimension drawings
Weight
Depending on device version
Material in contact with
media
Low-temperature carbon steel, stainless steel, Duplex steel
Electrical connection
Voltage
Flameproof pressurized electronics variant / electronics variant with terminal compartment in
increased type of protection:
Electrically isolated: 12 ... 24 V DC,
±10 %
Intrinsically safe electronics variant:
6 ... 16 V DC,
±10 %
PowerIn Technology™ with backup battery (2,400 mAh, 10.8 V),
optional for all electronics variants
Overvoltage category 1
Power input
Typically 0.45 W ... 2.45 W
Dependent on selected electronics configuration
Components fitted (optional)
Pressure and temperature
sensor
Values measured for pressure and temperature are used to correct the meter body geometry and to
determine the current Reynolds number.
Conversion method
PTZ (optional integrated)
Calculation method for
compressibility
SGERG88
AGA 8 Gross method 1
AGA 8 Gross method 2
AGA NX-19
AGA NX-19 mod.
NX-19 mod. (GOST)
GERG91 mod. (GOST)
Fixed value
GOST 3031.2-2015
Data archives
1 diagnostics archive (6,000 entries)
2 configurable measuring period archives (6,000 entries each)
Logbooks
Event logbook (1,000 entries)
Parameter logbook (200 entries)
Metrology logbook (50 entries)