2.9
Data processing in FLOWSIC600-XT
2.9.1
Logbooks
● Event logbook (1,000 entries)
Custody relevant and other events are written to the event logbook. It can store
1000 entries. As standard, no entries are overwritten when the logbook is full. An error
message is output.
The event logbook can be reset only when the parameter locking switch is open. Time
-
stamp, meter level, user ID of the active user, the event code and possibly supplemen
-
tary information are saved.
● Parameter logbook (200 entries)
The parameter logbook saves all parameter changes. It has a capacity of 200 entries
and is circulating by default. The oldest entries are deleted when the parameter logbook
overflows. Timestamp, meter level, user ID of active user, interface version, old value,
new value as well as the Modbus register number are saved.
When an overflowing/rolling logbook is set, the index numbers are further incremented
and the oldest entries overwritten with new data entries. Old entries are lost if no regular
backup with is made with the FLOWgate
TM
.
● Metrology logbook (50 entries)
Selected calibration-relevant parameters can be modified when the parameter locking
switch is closed and after logging in as authorized user.
An entry in the metrology logbook is generated to ensure the traceability of these
parameter changes. The following parameters can be changed when the parameter
locking switch is closed:
– Pulse valency
– Ambient pressure
– Minimum and maximum measurement pressure
– Default values for pressure and temperature
– Deactivation of Metrology Logbook
Timestamp, meter level, user ID of active user, old value, new value as well as the
Modbus register number are saved. The metrology logbook has a capacity of 50 entries
and stops by default when it is full. When the metrology logbook is full, changes of
custody-relevant parameters can only be carried out when the parameter locking switch
is open. The metrology logbook can be reset only when the parameter locking switch is
open.
NOTICE:
Type approval
Meter status “Measurement invalid” is activated when the FLOWSIC600-XT is
configured as a verifiable meter and the maximum number of entries is
reached in the metrology logbook or parameter logbook. Measured volumes
are registered in the malfunction volume encoder totalizers.