26
8024493/AE00/V1-0/2020-09| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-2K
Subject to change without notice
3
PRODUCT DESCRIPTION
3.3.4
MCUDH Ex-3K control unit
The control unit serves as user interface for the DUSTHUNTER SP100 Ex-2K sender/
receiver unit, prepares and outputs the measured values and also performs control and
monitoring functions.
In detail, the control unit takes over the following tasks, for example:
●
Sender/receiver unit power supply.
●
Output of measured values, computed data and operating states
●
Communication with the peripheral equipment
●
Output of error messages and other status signals.
●
Control of automatic test functions and access during service (diagnosis)
Via the service interface, the device parameters can be set using a computer and an
operating program. The parameters are stored reliably even in the case of a power failure.
Use the control unit according to the intended areas of application (
tration DUSTHUNTER SP100 Ex-2K”, page 33
).
Fig. 8: MCUDH Ex-3K control unit with options
WARNING:
Risk of explosion when the USB service interface is used in an
Ex-atmosphere.
Operation of the USB connector can cause an explosion.
▸
Using the USB service interface in an Ex-atmosphere is prohibited.
▸
If necessary, set up an alternative RS485 service interface that can be led into the
safe area (
Display module
Interface module (option)
Cable and line inlets (2×M25; 5×M20)
Power supply unit
Fastening brackets
Enclosure
Processor board
Terminal input, power supply