105
8024493/AE00/V1-0/2020-09| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-2K
Subject to change without notice
MAINTENANCE
12
4
Loosen the connecting wires on the Interface module.
Fig. 51: Illustrations, exchanging the Interface module
5
Use a screwdriver to unlock hat rail lock (1) of the power supply unit and remove the
module. Note how the ribbon cable is laid on the module.
6
Disconnect ribbon cable connector (2).
7
Connect ribbon cable connector (2) to the new Interface module.
8
Plug the new power supply unit onto the hat rail.
9
Connect the connecting wires to the new module.
10
Close the MCUDH Ex-3K with the control cabinet key.
11
Put the complete device back into operation (switch on voltage and check measured and
status values).