Chapter
6
Operating Instructions
CLP 100 Bar Code Reader
Operation
20
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8 008 912/0000/21-08-2002
If the CLP 100 is connected to a PC, the current parameters can be changed by using the
CLP Setup software. The parameter set of the default setting is saved in the CLP Setup and
in the CLP 100 and can be re-activated at any time.
The complete default setting can be viewed by using the CLP Setup software and printed
out, if required:
1. In order to save the current settings: Select the S
AVE AS
menu item in the F
ILE
menu bar
and enter the file name.
2. Click on
in the toolbar.
The default setting is loaded and is displayed in the tab cards.
3. Click on
in the toolbar.
The F
ILE
dialog box is opened.
4. Enter the comment for the header of the output and click on OK.
The dialog box for the print settings is opened.
5. Enter the settings for printing and click on OK.
CLP Setup prints out the default settings.
6.3
Quick-Start
1. Carry out the electrical installation in accordance with
A PC does not
have to be connected if the CLP 100 is operated with the default setting.
2. Switch on the supply voltage.
The LEDs of the CLP 100 light up and are extinguished after the self-test has been
completed successfully.
3. Present the bar code sample from
Figure 6-1
to the CLP 100 at a distance of approx.
35 mm.
4. Start the reading pulse: Cover the optical path of the sensor (close the switch).
Both LEDs light up.
Parameter
Default setting
Active code type
Code 39
Code length
Free
Multiple reading
2
Number of codes
1
Reading direction
Both
Send start/stop characters
Not active
Check digit
None
Reading pulse generation
Active low (unchangable)
Switching output
Not active
Host interface
RS-232 (unchangable)
Protocol
Start characters STX, Stop characters ETX
Data transmission rate
9,600 Bd
Data format
8 data bits, no parity, 1 stop bit
Output instant
Immediately
No Read message
BR
Tab. 6-1: Default setting of the parameters of the CLP 100