FACTORY SETTING
16
Factory setting
Parameter
Factory setting
SP1
80% of probe length measured from end of probe
RP1
5 mm below SP1
OUT1
Q1_Hno
SP2
For 5-pin versions: 20% of the probe length measured from the end of the probe
For 8-pin versions: 60% of the probe length measured from the end of the probe
RP2
5 mm below SP2
OUT2
Q2_Hno
TYP2
Q2_PNP
SP3
40% of probe length measured from end of probe
RP3
5 mm below SP3
OUT3
Q3_Hno
SP4
20% of probe length measured from end of probe
RP4
5 mm below SP4
OU4
Q4_Hno
TYP3
Q3_PNP
TYP4
Q4_PNP
QAHigh
50 mm below start of probe
QALOW
10 mm above end of probe
QAPOL
QA_Nrm
QATYP
Auto
QAFAIL
3.5 mA
SimCur
SimOff
SimVol
SimOff
DspVal
Distan
Filter
Off
SimLev
SimOff
TrsHld
100
MaskZn
0 mm
MaskTr
50%
Mode
Pulse
CalSta
noCal
Probe/Type
Depending on probe type: Rod/Rope
MaxCol
Depending on measuring mode: HiSped = AnySped, HiAcc = 10 cm/s
MeasMd
HiSpd
CalRng
6,005 mm
FomSta
Inactive
Limit
90
Offset
0 mm
Unit
mm
Lock
Inactive