O P E R AT I N G I N S T R U C T I O N S | C F P C U B I C
8021989 / 2017-10-20 | SICK AG
Subject to change without notice
3 2
8
OPERATION
8.5.6
Locking the display with a password
1.
Log in to Exprt mode, see
2. Activate or deactivate the display lock using CONFIG / LOCK.
If the display lock is activated:
−
The display is locked after a timeout of 5 minutes has elapsed
−
The display is unlocked temporarily by entering the password 000237
Note:
When the display is locked, only the configured measured value display can be seen.
8.5.7
Setting the medium adjustment
This setting is used to configure the speed of the medium adjustment. The medium
adjustment allows the sensor to adapt to changing measuring media. The slower this
adjustment occurs, the more stable the measurement.
Note:
The value is automatically set to 60 s when AUTCAL is performed.
Configuration
1.
Log in to Exprt mode, see
2. Set the parameter in the MEAS / ADAPT menu.
The possible values are 60 s, 30 s, 10 s, 5 s, 1 s.
If the medium remains the same, a value of 60 s is recommended.
8.5.8
Filtering measured values
Activating filtering
The filter supports the smoothing of the measured value; e.g., in the case of ripples on
level surfaces. For fast level changes, the average of the measured values over x sec
-
onds is indicated.
1.
Log in to Exprt mode, see
2. Set the parameter in the MEAS / FILTER menu.
The possible values are Off, 500 ms, 1 s, 2 s, 5 s, 10 s.
8.5.9
Testing the configuration
Testing outputs
Switching and analog outputs can be simulated. This allows for the wiring and signal
values on the connected systems, such as a PLC, to be checked.