O P E R AT I N G I N S T R U C T I O N S | C F P C U B I C
8021989 / 2017-10-20 | SICK AG
Subject to change without notice
3 0
8
OPERATION
•
Invert = Analog output signal is inverted; QaHigh 0 V and QaLow 10 V
6. Select electrical signal.
Set the parameter in the QA-### / QATYP menu to
0–10 V
.
8.4.4
Behavior of outputs in the event of an error
Configuration
Note:
The analog output designation (QA in the example here) changes according to the se-
lected process variable to be output.
This function is only available if 4 to 20 mA is selected under QA-TYP.
1.
Log in to Exprt mode, see
2. Select electrical signal.
•
Select the parameter in the QA-### / QA-TYP menu and set the value to 4 to
20 mA.
3.
Define the signal in the event of an error.
•
Select the parameter in the QA-### / QA-FAIL menu and set to the required
value.
8.5
Advanced functions
8.5.1
Selecting display unit
The following units can be selected for the measured values:
−
UnitLv: mm, inch
−
UnitTm: °C, °F
Configuration
1.
Log in to Exprt mode, see
2. Set the unit for level.
•
Set the unit in the CONFIG / UNITLV menu (mm/inch).
Or
1.
Log in to Exprt mode, see
2. Set the unit for temperature.
•
Set the unit in the CONFIG / UNITTM menu (°C/°F).
8.5.2
Setting an alternating display of measurands (DispA and DispB)
Up to two measured values can be displayed alternately.
Function
Description
Level
Level in mm/inch
Level %
Level in %
Temp
1)
Temperature in °C/°F
QaSign
1)
Qa analog output signal in mA/mV