NOTE
The following functions are deactivated in
Live
mode:
•
Digital inputs and outputs
•
Data output via the host interface.
2.
Align the device in the desired depth of field range with a medium-height object
with a test code.
3.
Click the
Camera & Illumination
configuration bar. Use the
Shutter timer
and
Brightness
sliders to adjust the image brightness so that the code is easy to see.
4.
Only available in
Extended
mode: activate the sharpness diagnostic bar. To do this,
go to the
Camera & Illumination
area and click the
Display sharpness
checkbox.
Variants with a compact C-mount lens: adjusting the brightness and sharpness
1.
Remove the optics protective hood. To do this, turn the optics protective hood
counter-clockwise as seen from the front, then detach and remove it,
bling the basic device with compact C-mount lens", page 28
2.
Undo the locking screw on the focus ring of the lens.
3.
Adjust the focus using the focus ring on the top side of the lens so that the online
image shows a sharp, clear image of the test code with no distortion.
✓
The test code in the image comes into focus and the edges are clearly discernible.
✓
Available in
Extended
mode: The sharpness diagnostic bar is now at its maximum
position.
4.
If necessary, use the
Shutter time
,
Brightness
and
Contrast
sliders to optimize the
brightness and contrast.
5.
Use the lock nut fitting to fix the sharpness ring setting in place.
6.
Attach the optics protective hood again and screw it tight.
Variant with C-mount lens and external ring illumination unit: adjusting the brightness
and sharpness
1.
If already fitted, remove the external ICL ring illumination unit. To do this, first
remove the connecting cable at the ring illumination unit. Undo the ICL ring illumi‐
nation unit at both mounting brackets and remove,
device with C-mount lens and external ICL ring illumination unit", page 29
.
2.
Remove the optics protective hood. To do this, turn the optics protective hood anti‐
clockwise as seen from the front then detach and remove it.
3.
Mount and connect the external ICL ring illumination unit again.
4.
Select and activate the fitted ICL ring illumination unit SOPAS ET:
°
Select the ICL illumination used. Path: SOPASET >
Camera & illumination
>
Illu‐
mination
>
External light connection
°
Activate the
External illumination
digital output. Path: SOPAS ET >
Interfaces &
digital outputs
>
Digital outputs
>
Output_Result 2
5.
Undo the locking screws on the aperture ring and focus ring of the lens.
6.
Adjust the aperture using the aperture ring (top ring) on the lens to a low value
(e.g., “2”).
7.
Reduce the
Shutter time
and
Brightness
parameters in SOPAS ET until the test code
is clearly visible in the image.
8.
Adjust the focus using the focus ring (lower ring) on the top side of the lens so that
the online image shows a sharp, clear image of the test code with no distortion.
✓
The test code in the image comes into focus and the edges are clearly discernible.
✓
Available in Extended mode: The sharpness diagnostic bar is now at its maximum
position.
9.
If necessary, use the
Shutter time
,
Brightness
and
Contrast
sliders to optimize the
brightness and contrast.
10. Use the lock nut fitting to fix the sharpness ring setting in place.
7
COMMISSIONING
54