7−4
7.3 Changing AC Power Supply Voltage
Follow the steps below to switch the AC power supply voltage
1.Turn off and disconnect the power cord and input/output cable from the amplifier unit
2.Remove the cover using two screws (M3) on the top face (Fig.7-6).
3.The AC power supply selection switch is positioned at the location shown in Fig.7-7. Selection
to 200 VAC, OFF, and 100 VAC is available. Voltage switching is available by sliding the switch
to the target voltage position. The fuse can support both 100 VAC and 200 VAC. The installed
fuse (100 VAC/200 VAC, 100 mA: 0334-3006) can therefore be used for both supply voltages.
4.Attach the amplifier cover so that the slit of the amplifier cover matches to the frame of the
amplifier unit.
5.Fasten the screws of upper face.
6.When using 200 VAC power supply, use optional AC power cord 200 V (Fig.7-8: 0311-5112).
OFF
100V
200V
Fig.7-8 AC Power cord for 200V
Front Panel
Rear Panel
Fig.7-6
Fig.7-7
-Power cord and input/output cable should be disconnected.
-As the AC power cord 200 V (0311-5112) has bare wire at one end,
processing is needed to connect to the power source.
-After switching the power supply voltage, change the power supply voltage
rating indicated on the plate on the amplifier cover.
(Length: 3.5 m, bare wire)
WARNING
Summary of Contents for 5683
Page 60: ...9 5 9 3 Dimensional Outline Drawing 9 3 1 Unit ...
Page 61: ...9 6 9 3 2 Panel Cut Sizes Square Hole ...
Page 63: ...9 8 9 3 4 Rack mounting Case Rack Rail Elongate Hole ...
Page 64: ...9 9 9 3 5 Fan Unit Fan 8CH Rack mounting Case Fan Unit Case ...
Page 65: ...9 10 9 3 6 Bridge Box 5370 and 5373 Attachment Hole ...
Page 66: ...9 11 9 3 7 Compact Bridge Box 5379 and 5380 ...
Page 67: ......