![Showa 5683 Operation Manual Download Page 22](http://html1.mh-extra.com/html/showa/5683/5683_operation-manual_1259939022.webp)
3−1
3
.
BEFORE MEASURING
3.1 Cable Connections
3.1.1 Input Cable Connections
(1) Paste a strain gauge to the location where measurement is made.
(2) Connect the strain gauge to the bridge box.
(3) Connect the bridge box or a transducer to the input connector
1
on the rear panel. For
connection information, see Cautions before Measuring on page 4-1. Since
Models
5683/5684
automatically adjusts the voltage drop generated between the bridge and the
amplifiers through the cable length compensation function, high-accuracy measurement is
possible. For more information on this adjustment function, see page 3-7.
(4) If you must measure large strains or wish to reduce the voltage going entering into the
transducer, you can change the bridge voltage to 0.5 V. Set the bridge voltage to 0.5 V using
bridge voltage selection switch 3 on the rear panel.
Fig.3-1 Connection with bridge box and so on
Bridge Box
When you yank this plug out of the
connector, you must hold this part
without rotating it.
Rear Panel
Input Connector
Use a strain gauge with short wires between the
measuring point and the bridge box in order to
reduce noise as much as possible. Do not cut or
add wire as the strain
gauge performs gauge
factor calibration with the wires connected.
Mark
Summary of Contents for 5683
Page 60: ...9 5 9 3 Dimensional Outline Drawing 9 3 1 Unit ...
Page 61: ...9 6 9 3 2 Panel Cut Sizes Square Hole ...
Page 63: ...9 8 9 3 4 Rack mounting Case Rack Rail Elongate Hole ...
Page 64: ...9 9 9 3 5 Fan Unit Fan 8CH Rack mounting Case Fan Unit Case ...
Page 65: ...9 10 9 3 6 Bridge Box 5370 and 5373 Attachment Hole ...
Page 66: ...9 11 9 3 7 Compact Bridge Box 5379 and 5380 ...
Page 67: ......