{11}
Shor-Line | 800.444.1579 |
www.shor-line.com
;
Section 4: Functions
•
power Button:
Provides power to the Warming Pad controller
and selects the time cycle simply by the number of
times pressed.
•
Press once: Will set the time for 60 minutes and the light
above ‘60 min’ on the control box will illuminate green.
•
Press twice: Will set the time for 90 minutes and the light
above ‘90 min’ on the control box will illuminate green.
•
Press three times: Will set the time for 120 minutes and the
light above ‘120 min’ on the control box will illuminate
green.
•
Press four times: Will turn the controller off.
•
Stand By:
During the last five minutes in any of the three
available time cycles the following will occur: For 4 minutes
the Stand By light will illuminate a solid yellow and a single
beep will occur every 30 seconds for a total of 8 beeps.
For the last minute of any time cycle the Stand By light will
illuminate a flashing yellow and a double beep will occur every
10 seconds.
•
alert:
The Warming Pad is factory set at 93° F (33.8° C) with
an over temperature protection of 97° F (36.11° C).
If for any reason the pad reaches a temperature of 97° F
(36.11° C), the heater will be disabled, the alert light
will illuminate red and flash continuously and a triple beep
will occur. This condition will remain until the pad lowers
in temperature or the unit is turned off by pressing the
Power Button.
•
NOTE:
The Warming Pad is factory set to reach a temperature of 93° F (33.8° C) which takes approximately
15 minutes. Turning the unit off does not immediately reduce the temperature. Should the alert condition occur
the patient should be removed from the Warming Pad as soon as possible to prevent possible burns.
•
Electrical Rating: 120/240 VAC, 50/60 HZ, .5A at 120 VAC, .25A at 240 VAC