• Time signal output setting items [When Time signal output (017) is selected]
Character
Setting Item, Function, Setting Range
Factory Default
Time signal output step
1
• Sets step number for time signal output performance.
• Setting range: 1 to 15
Time signal output OFF time
00:00
• Sets the Time signal output OFF time.
• Setting range: 00:00 to 99:59 (Time unit follows the selection in
[Step time unit] in the Program group.)
Time signal output ON time
00:00
• Sets the Time signal output ON time.
• Setting range: 00:00 to 99:59 (Time unit follows the selection in
[Step time unit] in the Program group.)
Time signal output
Time signal output activates during Time signal output ON time within the set step for
which Time signal output is performed.
Time signal output ON time follows Time signal output OFF time after the program
control starts.
The following program pattern shows that the temperature rises to 200 for 1 hour,
and stays at 200 for 2 hours after program control starts.
Step
1
2
Step SV
200
200
Step time
1:00
2:00
Time signal output (Fig. 6.3.4-5) is shown when set as follows.
•
The
step for which Time signal output is performed: 2
• Time signal output OFF time setting: 0:30
•
Time signal output ON time setting: 1:00
Step
1
2
Time signal output
(Fig. 6.3.4-5)
Time signal output is effective within the step set in [Time signal output step].
For example, if Time signal output ON time is set to “2:00” at the above,
Time signal output is turned OFF when step 2 is completed.
200
0
1:00
2:00
Program control run
ON
OFF
0:30 1:00
OFF time ON time