OPERATION (CONTINUED)
End of procedure
Verifying the Temperature Display Accuracy
1. Once the chamber temperature has stabilized, compare the reference
device and the oven display temperature readings.
a. If the readings are the same, or the difference between the two (2)
falls within the acceptable range of your protocol, the display is
accurately showing the oven chamber air temperature.
The
Temperature Verification procedure is now complete
.
b. See step 2 if a difference falls outside the acceptable range of your
protocol.
Reference Device
2. Perform a
temperature calibration
to match the controller display to
the actual chamber temperature if a difference falls outside your
protocol range.
a. Please see 36 in the User Maintenance section.
Reference Device