74
7 Appendix
AL4
Type:
Warning of gas concentration in ppm range
Adjustable:
Yes
Latching:
No
Trigger:
Alarm threshold AL4 exceeded
Indicator:
– Audible signal
– Visual signal
–
AL4 notification on display
Acknowledge-
ment:
– Possible for audible signal when alarm threshold
AL4 is exceeded
Reset:
– Automatically when level falls below alarm threshold
AL4
STEL
Type:
Main alarm (Short Time Exposure Limit)
Adjustable:
No
Latching:
Yes
Trigger:
Sum of the concentrations of a gas is greater than
the product of the OEL and the excess factor over the
averaging time
Indicator:
– Audible signal
– Visual signal
–
STEL
notification on display
Acknowledge-
ment:
– Not possible
Reset:
– By changing to another application
–
By switching device off
LTEL
Type:
Main alarm (Long Time Exposure Limit)
Adjustable:
No
Latching:
Yes
Trigger:
Sum of the concentrations of a gas is greater than the
OEL over the averaging time
Indicator:
– Audible signal
– Visual signal
–
LTEL
notification on display
Acknowledge-
ment:
– Not possible
Reset:
– By changing to another application
–
By switching device off