69
5 Maintenance
5.2.6.2 Adjusting the sensitivity for oxygen
The sensitivity for oxygen is adjusted with fresh air.
1. Make sure that only fresh air is being drawn in.
2. Open
Settings
.
3. Select
Adjustment
menu item.
4. Select
Adjustment O2
menu item.
5. Wait until the displayed reading is stable. (The reading may
still flash.)
6. Select
20.9 % VOL. (fresh air)
from the menu (select and
confirm with
OK
). This adjusts the sensitivity. The reading
shows 20.9 % vol.
5.3
Servicing
The device must only be serviced and repaired by SEWERIN
Service.
●
Send the device to SEWERIN for repairs and for annual main-
tenance.
Note:
If there is a service agreement in place, the device can be ser-
viced by the mobile maintenance service.
The inspection plate on the device shows con-
firmation of the last maintenance and the next
scheduled maintenance.
Fig. 21: Inspection plate