58
5 Maintenance
The device inspection is saved as a protocol. An overview with
the device inspection results is displayed.
This overview includes a list of all gas types for which the de-
vice is configured. Gas types for which the indication accuracy
was successfully tested in the device inspection are flagged
with
OK
. Gas types that are available but have not been tested
are flagged with
----
.
4.
Confirm the overview by pressing
OK
. The device returns to
measuring mode.
5.1.3
Testing the general status
The general status test is part of the device inspection (see sec-
). It is based on estimations by the user. The following
must be tested:
●
Housing
●
Signals
●
Probe
●
Filter
●
Pump
The battery charge status and the working condition of the controls
are automatically tested during the integrated device inspection.
The device inspection has been opened.
1. Select
General status
from the
Dev. Test ...
menu.
2. Test all associated subitems as described in section 5.1.3.1
3.
Confirm the prompt
General status OK?
by pressing
Yes
if
all
subitems show no faults during testing.
General status
OK
appears on the display.
This concludes the
General status
test.
5.1.3.1 Housing
●
Is the housing free from external damage?