EXPERT SYSTEM TEST (For Model 269)
1. Select UUT Test Tab.
2. Select “Auto-Detect” from UUT ID drop down box. Wait for software prompt to plug
EPIC cable into Transducer.
3. Remove Process Head from pressure sensor and plug in EPIC connected to MicroCal
top panel header. FIRST TIME UUT AUTO-DETECT -The MicroCal will automatically
download and save a new UUT Profile from a Setra 269.
A) UUT ID Selection Screen will appear.
B) Enter tag number or UUT ID number.
C) Tap OK.
D) UUT Part Number Screen will appear. Complete part # and tap OK.
E) All setup data for UUT will be automatically downloaded and stored under
new UUT ID Number.
F) Ready to initiate test.
4. Select Test Tab.
5. Press ‘Do Test’ and wait for completion
A) If “As Found” data is within the user-selected Re-Cal Accuracy- The MicroCal
saves the test data as both “As Found” and “As Left” and the calibration/testing is
complete.
B) If “As Found” data is outside the user-selected Re-Cal Accuracy- The MicroCal
initiates automatic zero and span adjustments (Auto Re-Cal). The MicroCal will
perform the Re-Cal process up to three times to meet the designated Re-Cal
Accuracy. One correction is usually sufficient to Re-Cal a UUT; however, if Re-Cal
is not achieved, then the MicroCal displays the following message: “UNABLE TO
ADJUST UNIT TO SPECIFIED RE-CAL ACCURACY”
C) Any Re-Cal process will result in test data being saved in ‘As Found’ and ‘As
Left’ data files.”
6. Remove EPIC Connector from UUT, and replace UUT Process Head to return UUT to
regular operation.