32
Diagram 19
Display Trend Graph on
UUT Warm-up
Display Test Results
Excitation Setting
Warning
Miswire Warning
Test Results
Save Warning
Manual Step-Through Test
Extra Digit Resolution
Restore Factory
Defaults
Maximum Leak Rate Input
Opens AUTO-DETECT
Default Settings Page
EDIT DEFAULTS PAGE
DISPLAY TREND GRAPH ON UUT WARM-UP: will make a graph appear when the test
page is opened, or the UUT ID is changed on test page. The graph will show UUT output
vs. time to allow the operator to see if the UUT has a warm-up drift, and output has
stabilized. (This feature is not likely to be required if the units to be tested are normally
connected to a power supply.)
DISPLAY TEST RESULTS IN GRAPH: will make a graph appear at the end of the automat-
ed test. Graph will show the UUT output error seen during test vs. applied pressure.
TURN ON EXCITATION WARNING SETTING: will make a dialog box appear on screen
anytime the excitation is about to be changed to a higher level.
(This is used to help prevent damage to unit. If all units to be tested can accept 24VDC
excitation, feature will not likely be required.)
TURN ON MISWIRE WARNING: will make a dialog box pop up when it appears the UUT
wiring or plumbing has been reversed, or if the UUT pressure range appears to be set
incorrectly.
TURN ON TEST RESULTS SAVE WARNING: will make a dialog box appear at any point
where unsaved test results could be deleted.
STOP AT EACH TEST POINT: Manual step-through test will stop a test in progress at each
pressure point and require user to press ‘OK’ on pop-up dialog.
ADD EXTRA DIGIT RESOLUTION: Add extra digit to pressure display increases pressure
resolution shown on real-time page. Adds one digit to the end.
RESTORE FACTORY SETTINGS resets settings to factory programming.
EDIT AUTO-DETECT Settings opens a page to access settings for AUTO-DETECT units
(269). (See Diagram 21)