30
6.6 SYSTEM
Screen Instructions (Diagram 18):
Data and system values:
Software Version - Version of software on MicroCal.
Firmware Version - Version of firmware on calibrator microprocessor board.
Serial Number - Serial number of device.
Calibration Date - Date of last calibration.
Filter Length - The number of samples taken when a data point is collected during an
automated test. It is also the number of samples taken when HOLD SAMPLE button is
pressed on Real Time page. The software will read this number of samples and return
the average as the data value. Maximum value will be 100.
System Volume - This variable cannot be set by the user. The value shown is fixed system
volume excluding transducer.
Excitation - This value is the current excitation being provided to +EXC.
Battery Level - This value is the current battery voltage level. Shows adapter power level
when 24V adapter is connected.
Edit Defaults - button, brings up default settings page. (see Diagram 19, pg. 32)
Set Engineering Units - Opens the Engineering Units Selection Page (for more, see
Diagram 21)