18
6.3 RUN TEST: GENERAL
Screen Instructions (Diagram 10):
1. Tap the RUN TEST tab at the bottom of the page.
2. Select desired UUT from the drop down box at the top of the page. If desired UUT is not
there, see Page 24 for creating a new UUT.
3. View the TEST drop down box and confirm that profile selected is the desired pressure
test profile, if not, select desired profile from drop down list or if desired test profile is
not there, see Page 22 for creating a new test setup.
4. Select AS FOUND or AS LEFT radio button.
5. Tap TARE and wait until completed. This zeroes the ref. unit.
6. Tap RUN button.
7. To stop a test in process, tap the STOP button. This will stop and
cancel the test.
8. Target pressure will be generated and data will be automatically
recorded at each pressure point.
Note: If the Display Test Results in Graph Option is selected (see
Edit Defaults page) - a graph will appear displaying pressure vs. error after the pressure sequence
has been completed (Diagram 11).
9. To record data - Tap RECORD TEST RESULTS button
Note: Record Test Results Button has 3 states indicating the save status of test results. When no test
results are available to save, the button reads “No Results To Save.” When results on screen have
been saved, the button will change to “Test Results Saved.” After a test is complete and before the
data is saved, the button will read “Record Test Results.”
10. To review data - Tap REVIEW TEST RESULTS button
11. To expand/collapse the Test Profile/Results window, double tap anywhere inside Test Results
list.
Diagram 10
Record Data
Target Pressure
Profile ID
Test Tab
Review Data
Do Test
Unit Under Test
(UUT) Identity
Re-Zero
Stop
As Found/As Left
Test Selection
Test Profile/
Results Window