48
11.2 INTE PRESSURE TRANSDUCER SPECIFICATIONS
Expressed in %FS as constant temperature.
Accuracy as RSS non-linearity, hysteresis and non-repeatability
ROOM SUM SQUARES (RSS)
NON-LINEARITY (±0.1%)
2
0.01%
HYSTERESIS (±0.05%)
2
0.0025%
NON-REPEATABILITY (±0.02%)
2
0.0004%
0.0129%
√0.0129%
±0.11%FS at constant temperature
Accuracy
(Worst Case) Over a Temperature Range of -10
o
F to 130
o
F
Non-Linearity:
±0.1% FS
Hysteresis:
±0.05% FS
Non-Repeatability:
±0.02% FS
Thermal Zero Shift:
<±0.32% FS
Thermal Span Shift:
<±0.24% FS
Zero Offset:
±0.2% FS
Span Offset:
±0.2% FS
±1.13% FS
Long-term stability error not included.
Total Error Band