SOL-96.110.071 / 180122
TOPAZ Aluminum
Maintenance
32
6.2. Stop of Operation for Maintenance
Before performing any maintenance work on reagent or pump tubes,
it is recommended to flush the reagent system with deionized water.
Note:
• Before flushing the reagent tubes, make sure that the sample
inlet tap is open.
• To avoid chemical reactions in the photometer, flush the
reagent tubes one after the other, not simultaneously.
Proceed as follows:
1
Remove all reagent tubes from their bottles and place them in a
beaker filled with deionized water.
2
Open the <PUMPS> tab.
3
Press the <ON> button next to “P1 - Buffer pH6-6.2”.
4
Let the pump run until it stops automatically.
This takes approximately 2.5 minutes.
5
Repeat steps 3–4 with pumps “P2 - Phenantroline”, “P3 - Pyro
-
cathecol Violet” and “P9 - HCl 2N + 400µg Al”.
ON
ON
ON
ON
OFF
OFF
OFF
OFF
P1 - Buffer
P1 - Buffer pH6-6.2
P2 - Phenantroline
P3 - Pyrocathecol Violet
P9 - HCl 2N + 400μg Al