SOL-96.110.071 / 180122
TOPAZ Aluminum
Maintenance
41
6.6.2
Zero Calibration
Note:
• On multi-channel instruments, only calibrate sample stream 1.
Procedure
The procedure is similar to the manual measurement procedure
. Use deionized water in
-
stead of standard.
Results
screen
At the end of the zero calibration, a window displays the previous
(actual) and the new (calculated) offset. Confirm by pressing the
button.
6.6.3
Standard Calibration
Note:
• On multi-channel instruments, only calibrate sample stream 1.
Procedure
The procedure is similar to the manual measurement procedure
.
Begin calibration by entering the reference value of the standard
used. Pay attention to the correct concentration unit (ppb by default).
Then follow the instructions on the screen.
Results
screen
At the end of the standard calibration, a window displays the previ
-
ous (actual) and the new (calculated) calibration coefficient and the
concentration measurement readings before and after calibration.
Confirm by pressing the button.
CALIBRATION ZERO RESULTS
Actual offset
Offset Calculated
-0.1
0.2
CALIBRATION RESULTS
Actual coefficient
Coef. Calculated
Measure before calib
Measure after calib
1.040
1.047
20.15
20.00