![Seres OL SOL-96.110.071 Operator'S Manual Download Page 28](http://html1.mh-extra.com/html/seres-ol/sol-96-110-071/sol-96-110-071_operators-manual_1246339028.webp)
SOL-96.110.071 / 180122
TOPAZ Aluminum
Startup
26
Check black
reagent tube
and stirrer
To check the flow through the black reagent tube and the proper
function of the stirrer, perform the following steps:
10
Open the <ACTUATOR> tab.
11
Set “SV Emptying” to <ON> and “SV Filling” to <OFF> and wait
until the instrument is empty.
12
Then set "SV Emptying” and “Stirrer” to <OFF> as well.
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Stirrer
SV Emptying
Heating
SV Filling
ACTUATOR
4-20mA OUTPUT
PUMPS
RELAYS
INPUTS
Vessel temperature :
27.48 °C
1.
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Stirrer
SV Emptying
Heating
SV Filling
ACTUATOR
4-20mA OUTPUT
PUMPS
RELAYS
INPUTS
Vessel temperature :
27.48 °C
2.