SOL-96.110.071 / 180122
TOPAZ Aluminum
Troubleshooting
57
42
Division by zero -
Absorbance = 0 or Current
Calibration Coeff. = 0
– Check if the calibration coefficient is
equal to zero. In this case, perform a
new calibration.
– Check if the final absorbance is equal
to zero. If this is the case, the reaction
has not occurred.
Perform the following steps:
– Check that the reagents are well
injected into the photometer.
– Check that the stirrer turns.
43
Division by zero - Absorbance =
-999998.0
– Check that the optical fibers are well
connected to their holder on the door.
44
Value out of range
– Check that the analyzer is well
calibrated, check standard solution
concentration by a manual measure
cycle.
45
Division by zero - Absorbance =
-999998.0
– Check that the optical fibers are well
connected to their holder on the door.
46
Incorrect projector power
– Contact customer service
47
Threshold 1 exceeded
– Check process
– Check programmed value for
threshold 1
48
Threshold 2 exceeded
– Check process
– Check programmed value for
threshold 2
49
Optical setting incorrect
(analyzer cannot adjust the
optical gain)
– Check that the optical fibers are well
connected to their holder on the door.
– Check that the photometer is not dirty.
– Check that the analyzer is supplied
with water.
– Check that the stirrer turns.
No.
Text shown on display
Corrective action