Sentry Equipment Corp

4. 

Reassemble the sampler and clamp-mount onto the piping tee .

Starting the Sampler

 

Â

NOTE

Delayed start is designed to avoid sampling product which may be trapped in a transfer hose or other piping at the 

conclusion of pumping a load . Delay before the first sample should be adjusted to the minimum time needed to clear the 

hose of standing product and ensure that product at the sampler is from load being pumped, not carryover from a previous 

producer .

Delay time adjustment is made by changing the setting on the delay potentiometer located in the upper left-hand corner of 

the timing module .

Turn off power to the sampler controller before opening the enclosure or making internal adjustments .

1. 

With the power OFF and the controller enclosure open, locate the delay potentiometer on the timer module . The 

dial is calibrated in percentages [%] of 0-100 . Adjust by turning the knob with a screwdriver in the slot provided .

2. 

Start delay adjustment (0 .1 - 10 seconds) .

3. 

Turn clockwise to increase, counterclockwise to decrease 

delay .

4. 

Start delay is a single turn potentiometer . It will not turn more 

than a nominal 270° . Adjustment is 1/10 second minimum to 

10 seconds maximum .

5. 

When the sampler controller is operated through a DELAY-

SAMPLE OFF sequence, allow approximately 10 seconds for 

the delay timer to reset before starting the next sampling 

operation .

6. 

After adjusting the delay time, replace the controller cover 

securely onto the box .

Maintenance

Refer to your specific group drawing that is included with this manual .

The ISOLOK MSA sampler should be cleaned and checked at daily and weekly intervals for proper operation and to 

ensure any wear is detected so preventive maintenance can be performed .

Never use pipe wrenches, vise-grip pliers or other marking or abrasive tools on your sampler . 

Daily Cleaning

Clean the sampler by removing the unit from the line tee, partially disassemble it, and manually clean it . Typical steps 

for cleaning the sampler include:

50

100

0

Summary of Contents for ISOLOK MSA

Page 1: ...Original Instructions Installation Operation Maintenance Manual ISOLOK MSA Sampler Point Samplers S AS IOM 00300 8 10 22...

Page 2: ...ability Sentry Equipment Corp its employees agents and the authors and contributors to this document specifically disclaim all liabilities and warranties express or implied including warranties of mer...

Page 3: ...Controller Installation 8 Inspection 11 Operation 11 Initial Cleaning 11 Starting the Sampler 11 Maintenance 12 Daily Cleaning 12 Weekly Inspection 13 Complete Sampler Disassembly and Maintenance 13 C...

Page 4: ...ons DANGER indicates a hazardous situation which if not avoided will result in death or serious injury WARNING indicates a hazardous situation which if not avoided could result in death or serious inj...

Page 5: ...at the equipment is isolated from all connecting piping the equipment is depressurized the contents have been drained and the equipment is cool Moving parts This equipment may contain moving parts All...

Page 6: ...se or piping to pass the sampler before operation begins During initial delay the controller solenoid valve is energized causing the sampler plunger to extend into the mounting tee where it can be flu...

Page 7: ...y plastic bags that close with a built in wire tie Sizes up to about an 8 oz 250 ml capacity can be supported by carefully folding the open bag mouth and wire around the small diameter below the drip...

Page 8: ...sampler controller in its most effective location 1 The line where the sampler is to be located can be horizontal vertical or at any angle The location selected must mount the sampler body with its l...

Page 9: ...ed around a sealing bulkhead wall without the danger of pinching the tube or blocking the door Wall Penetrations The following procedure is suggested whenever air tube must be routed through a wall be...

Page 10: ...miting device This device automatically disconnects power to the controller in the event of an accidental overload or short circuit Identified as AutoFuse this device opens the power circuit any time...

Page 11: ...After installation the sampler periodically should be checked for proper operation and to ensure that any wear is detected so proper preventive maintenance can be performed The Maintenance section of...

Page 12: ...knob with a screwdriver in the slot provided 2 Start delay adjustment 0 1 10 seconds 3 Turn clockwise to increase counterclockwise to decrease delay 4 Start delay is a single turn potentiometer It wi...

Page 13: ...r body to compress the spring Install the clamp Replace any port adapter including the O ring seal 10 Replace the line tee gasket and install the sampler and clamp in place The sampler body should be...

Page 14: ...mble the four plunger seals rod seal and piston seal O rings onto the plunger 2 Apply a thin film of the approved dairy lubricant to all six seals Assemble the spring onto the plunger and insert it in...

Page 15: ...roller sample rate potentiometer is a round component with brown wire leads on two of three lugs If controller operation becomes erratic e g cycles at some load size dial settings but not at others or...

Page 16: ...alve using the following steps 1 Disconnect the air line tubing to the sampler Disconnect the incoming compressed air supply line Open the sampler controller enclosure Disconnect the wire leads betwee...

Page 17: ...y monitor and measure processes for improved production efficiency output and safety Standing behind our commitments we are determined to tackle any application anywhere We know that running an effici...

Page 18: ...r combining both into one connection and using the remaining pin for the sampler power lead AC Power Supply Package This parts package is used with 230 VAC supply and rectifies and transforms 230 VAC...

Page 19: ...ation NOTE Within the motor junction box either T5 or T8 normally will be connected to T2 and T3 Select a combination of motor leads so that the sampler is NOT wired into this connection Polarity The...

Page 20: ...d U S Dept of Transportation or equal for service to at least 120 psi 827 KPa Tanks can be obtained locally from firms that build or repair truck air brake systems or an air tank package is available...

Page 21: ...roduces less than 1 U S ounce 30 cc or 60 sampler cycles A biased sample could result 3 Select a sample container size that will not be filled completely even on longest sampling times Use care in set...

Page 22: ...inute etc actual delivery will be influenced by many factors some of which are Speed of pump drive and pump condition Type of valve on farm tank Suction hose size and condition Head pressure suction a...

Page 23: ...ISOLOK MSA Sampler 23 This page is intentionally left blank...

Page 24: ...sentry equip com 966 Blue Ribbon Circle North Oconomowoc WI 53066 U S A 1 262 567 7256 support sentry equip com Serving customers in more than 50 countries across six continents worldwide...

Reviews:

Related manuals for ISOLOK MSA