background image

ISOLOK MSA Sampler 11

Air Supply Pressure

The sampler is designed to operate best at a regulated pressure of 40 psi (275 KPa) . The regulator inside the controller 

has been factory set to provide 40 psi to the sampler and should not require field adjustment . However, if problems 

occur and are attributable to variation in the air pressure to the sampler, regulated pressure can be checked and 

readjusted as follows:

1. 

Remove the 1/8 in plug from the side of the regulator body . Install an accurate pressure gauge on this port .

2. 

With the air supply ON and sampler cycling every 2 seconds, the gauge should read 40 psi between cycles .

3. 

If the gauge does NOT read 40 psi, open the controller enclosure and readjust as follows:

a. 

Pull the adjusting knob upward to release the setting .

b. 

Turn the knob while the sampler is cycling until the gauge reads 40 psi between cycles . Clockwise [+] rotation 

increases, and counter-clockwise [-] rotation decreases the system pressure .

c. 

Push the adjusting knob downward to lock the pressure setting .

d. 

Turn the sampler controller OFF and disconnect the electrical power supply . Shut off the air supply . Remove 

the gauge, replace the plug and close the controller cover .

Inspection

After installation, the sampler periodically should be checked for proper operation and to ensure that any wear is 

detected so proper preventive maintenance can be performed . The Maintenance section of this manual describes 

recommended procedures fo daily, weekly and other cleaning and maintenance .

Start-up Inspection

With electrical power and compressed air ON, set the dial pointer to approximately 40 percent of the maximum rate 

(in the 10 o’clock position) . Remove the sampler from the line . Turn the controller switch to ON for a few seconds and 

observe sampler operations .

Each time the sampler cycles, the plunger will extend out of the body approximately 7/8 in” (22 mm), then withdraw . 

Each time the plunger moves in or out, you should hear a firm “click” as the piston bottoms at each end . The plunger 

dwells inside the sampler body between cycles .

A complete plunger stroke occurs in about 1/5 (0 .2) second . If plunger seals are dry, a sluggish or incomplete stroke 

may occur . See the Maintenance section of this manual for service information

Operation

Initial Cleaning

1. 

Remove the sampler body .

2. 

Using approved sanitary methods, clean the parts that have contact with product (i .e ., plunger rod, front face, 

interior and discharge ports of the sampler body) .

3. 

Wipe a very thin film of approved dairy lube (white grease, CIP lube or Vaseline) on the four plunger O-ring seals, 

plus the rod and piston O-rings .

Summary of Contents for ISOLOK MSA

Page 1: ...Original Instructions Installation Operation Maintenance Manual ISOLOK MSA Sampler Point Samplers S AS IOM 00300 8 10 22...

Page 2: ...ability Sentry Equipment Corp its employees agents and the authors and contributors to this document specifically disclaim all liabilities and warranties express or implied including warranties of mer...

Page 3: ...Controller Installation 8 Inspection 11 Operation 11 Initial Cleaning 11 Starting the Sampler 11 Maintenance 12 Daily Cleaning 12 Weekly Inspection 13 Complete Sampler Disassembly and Maintenance 13 C...

Page 4: ...ons DANGER indicates a hazardous situation which if not avoided will result in death or serious injury WARNING indicates a hazardous situation which if not avoided could result in death or serious inj...

Page 5: ...at the equipment is isolated from all connecting piping the equipment is depressurized the contents have been drained and the equipment is cool Moving parts This equipment may contain moving parts All...

Page 6: ...se or piping to pass the sampler before operation begins During initial delay the controller solenoid valve is energized causing the sampler plunger to extend into the mounting tee where it can be flu...

Page 7: ...y plastic bags that close with a built in wire tie Sizes up to about an 8 oz 250 ml capacity can be supported by carefully folding the open bag mouth and wire around the small diameter below the drip...

Page 8: ...sampler controller in its most effective location 1 The line where the sampler is to be located can be horizontal vertical or at any angle The location selected must mount the sampler body with its l...

Page 9: ...ed around a sealing bulkhead wall without the danger of pinching the tube or blocking the door Wall Penetrations The following procedure is suggested whenever air tube must be routed through a wall be...

Page 10: ...miting device This device automatically disconnects power to the controller in the event of an accidental overload or short circuit Identified as AutoFuse this device opens the power circuit any time...

Page 11: ...After installation the sampler periodically should be checked for proper operation and to ensure that any wear is detected so proper preventive maintenance can be performed The Maintenance section of...

Page 12: ...knob with a screwdriver in the slot provided 2 Start delay adjustment 0 1 10 seconds 3 Turn clockwise to increase counterclockwise to decrease delay 4 Start delay is a single turn potentiometer It wi...

Page 13: ...r body to compress the spring Install the clamp Replace any port adapter including the O ring seal 10 Replace the line tee gasket and install the sampler and clamp in place The sampler body should be...

Page 14: ...mble the four plunger seals rod seal and piston seal O rings onto the plunger 2 Apply a thin film of the approved dairy lubricant to all six seals Assemble the spring onto the plunger and insert it in...

Page 15: ...roller sample rate potentiometer is a round component with brown wire leads on two of three lugs If controller operation becomes erratic e g cycles at some load size dial settings but not at others or...

Page 16: ...alve using the following steps 1 Disconnect the air line tubing to the sampler Disconnect the incoming compressed air supply line Open the sampler controller enclosure Disconnect the wire leads betwee...

Page 17: ...y monitor and measure processes for improved production efficiency output and safety Standing behind our commitments we are determined to tackle any application anywhere We know that running an effici...

Page 18: ...r combining both into one connection and using the remaining pin for the sampler power lead AC Power Supply Package This parts package is used with 230 VAC supply and rectifies and transforms 230 VAC...

Page 19: ...ation NOTE Within the motor junction box either T5 or T8 normally will be connected to T2 and T3 Select a combination of motor leads so that the sampler is NOT wired into this connection Polarity The...

Page 20: ...d U S Dept of Transportation or equal for service to at least 120 psi 827 KPa Tanks can be obtained locally from firms that build or repair truck air brake systems or an air tank package is available...

Page 21: ...roduces less than 1 U S ounce 30 cc or 60 sampler cycles A biased sample could result 3 Select a sample container size that will not be filled completely even on longest sampling times Use care in set...

Page 22: ...inute etc actual delivery will be influenced by many factors some of which are Speed of pump drive and pump condition Type of valve on farm tank Suction hose size and condition Head pressure suction a...

Page 23: ...ISOLOK MSA Sampler 23 This page is intentionally left blank...

Page 24: ...sentry equip com 966 Blue Ribbon Circle North Oconomowoc WI 53066 U S A 1 262 567 7256 support sentry equip com Serving customers in more than 50 countries across six continents worldwide...

Reviews:

Related manuals for ISOLOK MSA