![Sentek RT6 Hardware Manual Download Page 28](http://html.mh-extra.com/html/sentek/rt6/rt6_hardware-manual_1244388028.webp)
SENTEK
Page 24
Power sources
Ni-Cad charger
Power modes
The following power sources can be applied to the
Main Board :
Solar panel which is rated at 10 watts at 12 volts
External DC supply which can range from 16 to
24 volts DC capable of supplying 1.5 amps. This
can be from available DC supply or via two 12
volt car batteries (in series to give 24 volts)
External AC supply which can range from 12 to
16 volts AC. This can be from available low
voltage AC supply or via a 240 volt AC plug
pack which delivers 16 volts AC at 1.5 amps.
The data logger board is fitted with a Ni-Cad battery
which is also charged by the solar panel (or external
power supply) or the SLA battery. This battery
protects the contents of the data memory and
time/date clock if the logger is removed to download
data in the office. Once removed from the logging
facility the data is secure for approximately 2 days.
Connecting the logger to an "Office Download
Cable" which is connected to power will charge this
battery, thus prolonging the time the logger can be
removed from the logging facility.
There are 3 power modes:
1.
No logger connected. With the battery cable
plugged into the main board and without the
logger plugged in, the battery charging circuit is
active and is ready to receive charge from the
external supply. This is done so that the battery
can continue being charged while the logger has
been removed from the logging facility for
downloading.
2.
Logger present. When the logger is connected to
the unit then power is applied only to the
circuitry that runs continuously. This is done to
minimize power drain.
3.
Sampling Sensors. During sampling of data from
the sensors, only the circuitry that is required is
powered. This is under control of the logger and
is done also to minimize power requirements of
the system.
Summary of Contents for RT6
Page 1: ...Sentek RT6 HARDWARE MANUAL Version 3 2 February 2011 ...
Page 2: ......
Page 4: ......
Page 8: ...SENTEK Page 4 ...
Page 30: ...SENTEK Page 26 ...
Page 46: ...SENTEK Page 42 ...
Page 50: ...SENTEK Page 46 ...
Page 62: ...SENTEK Page 58 ...
Page 63: ...APPENDIX A Page 59 APPENDIX A System Overview ...
Page 64: ...SENTEK Page 60 ...
Page 65: ...APPENDIX B Page 61 APPENDIX B System configuration examples ...
Page 66: ...SENTEK Page 62 ...
Page 67: ...APPENDIX C Page 63 APPENDIX C Probe configuration examples ...
Page 68: ...SENTEK Page 64 ...
Page 69: ...APPENDIX D Page 65 APPENDIX D Front panel visual ...
Page 70: ...SENTEK Page 66 ...
Page 71: ...APPENDIX E Page 67 APPENDIX E Overlay diagram RT6 Main Board Overlay ...
Page 72: ...SENTEK Page 68 ...
Page 73: ...APPENDIX F Page 69 APPENDIX F Cable Wiring Order 5 wire Connector ...
Page 74: ...SENTEK Page 70 ...
Page 78: ...SENTEK Page 74 ...
Page 80: ...SENTEK Page 76 ...
Page 82: ...SENTEK Page 78 ...