Cambridge Sensotec Rapidox 1100L Oxygen Analyser Instruction Manual
or ‘ppm’. To access a field, either click inside it or use the TAB key to scroll
through the fields. The field parameters are as follows:
8.5.1
Config 1 Screen
1)
Oxygen Range:
Not available for the Rapidox 1100L
2)
Oxygen Units:
See Section 7.8 for a full description. Use the drop down
box to select AUTO% or PPM. The default setting is AUTO%.
3)
Temperature Units:
See Section 7.8 for a full description. Use the drop
down box to select °C or °F.
4)
Pressure Units:
See sections 7.8 for a full description. Use the drop down
menu to select MBAR, BAR, PSI, TORR or KPa as the unit for pressure
measurement. The default setting is MBAR.
5)
Pressure Correction Mode:
See section 7.11 for a full description. Use the
drop down menu to select NONE or AUTOMATIC. The default setting is
AUTOMATIC.
6)
Operating Mode:
Not available for the Rapidox 1100L
7)
Auxiliary Output Parameter:
Use this menu to select either PRESSURE or
TEMPERATURE as the parameter sent to the auxiliary output on the rear
panel of the analyser.
8)
LCD Refresh interval:
Use the drop down menu to select a refresh rate for
the LCD (min 0.1, max 1.5 sec). The default is 0.3 seconds.
9)
LCD Second Line Option:
See section 7.9 for a full description. Use the
drop down menu to select Pressure, Temperature, None, Date/Time or
alternating Temperature and Pressure.
10)
Pump Flow Rate:
See section 7.3 for a full description. Use the drop down
menu to select a pump flow value from 10 to 100% or OFF.
11)
Password option:
See section 7.4 for a full description. The password is
factory disabled and set to ‘0000’. Use the drop down menu to enable or
disable the password feature and the password box ‘****’ can be used to
enter a new password. Note that you will be prompted to enter the original
password if you wish to make changes to the password or its status.
Version 1.6
21
Last printed 16/11/2011 16:59:00