(Options: Press. Temp.)
-
®
-
4-Set Outputs
(prompts 1)
_0.000E+00 = 0V
(prompts 3)
Edit O
2
(ppm)
1.000E+04 = 10V
(prompts 3)
Aux out: Press.
(prompts 6)
_0.000E+00 = 0V
(prompts 3)
1.000E+02 = 10V
Outputs set
Edit press(mbar) or Edit temp(°C)
Edit O
2
(ppm)
®
®
®
®
®
ENT
ENT
ENT
ENT
ENT
®
®
1.000E+02 = 10V
(prompts 3)
Edit press(mbar) or Edit temp(°C)
Displays message for 1.5 seconds
®
ENT
®
3-Set Alarms
(Prompts 1)
Set: Alarm 1
(Prompts 2)
Assign to: O
2
(Prompts 2)
(Options: O
2
, Temp., Press)
Outputs: On
(Prompts 6)
(Options: On, Off)
Audible: On
(Prompts 6)
Visible: On
(Prompts 6)
Polarity:
(Prompts 6)
(Options: On, Off)
(Options: On, Off)
(Options:
)
(Options: Alarm 1, Alarm 2)
1.000E+01
(Prompts 3)
Contacts: N/O
(Prompts 6)
®
®
®
®
®
®
®
®
®
®
Contacts: N/O
Alarm 1 set
ENT
®
®
®
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
Edit Alarm Setpoint
(Options: N/O, N/C)
Displays message for 1.5 seconds
ESC
ESC
®
®
®
®
ENT
1-Calibrate
(prompts 1)
4.000E+04ppm
(prompts 3)
Enter for two seconds to calibrate
Displays message for 1.5 seconds
Displays message for 1.5 seconds
Enter for two seconds to calibrate
APPENDIX 1: MENU FLOWCHART FOR RAPIDOX 1100L
20ppm
O
2
0.0mbar
ESC
Calib: O
2
High
(prompts 2)
ENT
Continues on page 2
®
Back to Menu List : Changes Applied
Calib: O
2
Low
(prompts 2)
ENT
ENT
0.000E+00ppm
(prompts 3)
4.000E+03ppm O2
Calib. O
2
High
0.000E+00ppm O2
Calib. O
2
Low
®
ENT
®
ENT
4.000E+03ppm O2
O
2
calibrated
0.000E+00ppm O2
O
2
Calibrated
®
®
®
®
®
ESC - Back to Menu List. No changes
Edit 0
2
High
Edit 0
2
Low
ESC
®
2-New Sensor
(prompts 1)
New Sensor?
(prompts 7)
New Sensor
Confirmed
Displays message for 1.5 seconds
®
ENT
Fit New Sensor
in 5 Days
Displays message for 1 second
Confirm
ENT
®
®
Back to Measurements : Changes Applied
-
®
®
®