Use the UP & DOWN arrows to select mbar, bar, torr, psi or kPa and press ENT.
The display will now show the pressure in the selected units.
7.11
Pressure Mode
The user has the option of using the reading from the internal pressure sensor to
automatically correct for changes in gas pressure and hence oxygen pressure. To
access this function scroll down to menu option 7 “Pressure Mode” and press
ENT. The desired mode can then be selected using the UP & DOWN arrows and
pressing ENT to store and proceed. The two modes available are:
1.
None
:
If this mode is selected then there is no pressure correction
performed on the oxygen reading. The pressure and oxygen sensor act
independently, and the oxygen reading will vary with the gas pressure.
2.
Auto
:
This is the factory default setting. If the Auto mode is selected then
the Rapidox 1100L uses the reading from the internal pressure sensor to
correct the oxygen partial pressure and maintain an accurate concentration
reading.
7.12
Setting the Baud Rate
If you are using the software for data logging you can choose the communications
speed (Baud Rate) by scrolling down the menu to option 8 “Set Baud Rate” and
using the UP & DOWN arrows to select 9600, 57,600 or 115,200. Press ENT
button again to store and return to normal operation. The default factory setting is
115,200. However if you are using an older PC you can decrease this to 57,600 or
even 9600 to improve the reliability of the software and data communications.
7.13
RS232 / RS485 Port
The data communications port is normally configured to RS232. It is also possible
to factory set the data communications port to RS485. Once configured at the
factory it is not possible to change unless the unit is returned to Cambridge
Sensotec Ltd.
Data from the Rapidox is sent on demand to the 9-way 'D' type socket (7) on the
rear panel and can be read using a simple terminal programme such as
'HyperTerminal' in Windows, or with the user's own custom software.
Alternatively a simple and convenient data logger programme is included with the
supplied software and is described in section 9 below.
7.13.1
RS232 Protocol
Rapidox data can be read from the RS232 port at the back of the unit. The default
RS232 configuration is 115200-8-N-1, as shown in this example for COM1: