Cambridge Sensotec Rapidox 1100L Oxygen Analyser Instruction Manual
Figure 17: Data Logging page
During logging, data from the analyser is shown in the central area of this page.
The data is displayed in the following columns: elapsed time (in seconds, minutes
or hours), the raw signal from the sensor amplifier (in mV), the oxygen value (in
ppm scientific notation), pressure in mbar, the temperature (in °C), the current
time, the date (in regional format), and status (displaying the word ‘ALMx’ if any
alarms are activated).
To start data logging select the frequency of measurement by entering a time and
selecting a unit from the drop down list next to the ‘Read every’ label. For
example if you wish to log data once every minute, type ‘1’ in the box and select
‘Mins’ from the drop down menu.
You can open the NOTES window at any time by selecting ‘Notes’ from the ‘Data
Logger’ drop down menu. A small text box appears in which you can type a text
message. This note is then saved as a header in the data file set and can be viewed
along with the data.
You can either choose to have data logged continuously until ‘Stop’ is selected
from the Data Logger menu, or you can choose to set a time limit for data logging
by checking the box labelled ‘Time Limit’, setting the duration in the ‘Run For’ box
and selecting units with the drop down list. For example if you wish to log data
for thirty minutes type ‘30’ in the box and select ‘Mins’ from the drop down list.