Cambridge Sensotec Rapidox 1100L Oxygen Analyser Instruction Manual
7
6
8 9
10
Figure 2: Rapidox 1100L rear panel. Numbers are referred to in the text.
7.1
The Rapidox Analyser - Menu System
All of the user-programmable functions are accessed via a menu system which is
controlled using the front panel Keypad (2). To access the menu press the ENT
button and to escape and return to the operating screen press ESC at any time.
The menu system flow chart is shown in Appendix 1 at the back of the manual.
7.2
Getting Started
Ensure that the Rapidox analyser is located away from extreme heat and dirt
environments. Plug the unit in to a suitable supply (noting the information on the
serial sticker and using the power cable supplied) using the rear power socket (8).
Make sure that the cooling fan (9) is not obstructed during operation.
Turn the unit on using the red power switch on the front (5). The LCD (1) will
display the serial number and firmware version for a few seconds before
measurements begin. The sensor will take approximately five minutes to stabilise.
The LCD display will show the oxygen reading and either the temperature,
pressure or date and time (user-selectable). The symbol AL1 or AL2 may appear
to the bottom right of the LCD if the alarm system is enabled and an alarm
condition applies.
Push the gas sample tube (6mm OD/4mm ID) into the GAS IN (3) quick release
connector. Repeat for the GAS OUT (4) connector should an outlet tube be
required. The pump will draw gas at a flow rate depending on the value set either
in the software or from the menu. The default setting is OFF.