Allow five minutes for the Rapidox to stabilise fully. If the sensor has been
exposed to air or the gas composition is greater than 3% oxygen then the message
“O/Range” will appear on the display. Below 3% oxygen but above 1% the display
will read the maximum value of 10500ppm (or 1.05%) but will flash until the gas
mix goes below 10500ppm, when normal readings will begin. Flush the analyser
with your measurement gas and wait a few moments for the readings to come on
scale.
7.3
Pump Control
The flow rate of gas drawn into the analyser can be controlled using the menu
option 2 “Set Pump Flow”. The unit is supplied with the flow rate set to OFF. If
you wish to change the flow rate press ENT and select menu option 2. Use the UP
& DOWN arrows to change the value from OFF to 100% in 10% increments. 100%
equates to approximately 1.5 litres per minute. Once you have selected the
desired flow rate, press ENT to confirm your selection. If you do not press ENT,
the pump will revert to the previous rate when the unit exits the menu system.
Alternatively the pump can be switched on or off using the switch (10) on the rear
panel. Please make sure that this is switched on and the pump is running. When
using the unit with a pressurised (flowing) gas source the pump can be left
switched off.
7.4
Menu Access / Passwords
The analyser has an option to set a password that will restrict access to the menus.
The password menu is disabled by default in the factory. If you wish to password
protect the analyser press ENT and scroll down to the PASSWORD option (menu
option 12). Pressing ENT again will ask you for the default password which is
“0000”. Enter this using the UP and DOWN arrows. Once the password has been
entered successfully you can choose between ENABLE and DISABLE in the menu
using the UP and DOWN keys.
The other option is to CHANGE the password to a new value. The password must
be 4 digits long and can be any combination of numbers from 0000-9999. Make
sure that the new password is noted down!
If the password function has been enabled then in the future each time the menu
system is accessed, the password prompt will display. After the correct password
is entered then the full menu will be accessible. You can make as many changes
as you want to the menu options and only press ESC once you have finished.
Pressing ESC takes you out of the password protected area and you will have to re-
enter the password to go back in again.
If the password has been forgotten please contact Cambridge Sensotec who will
advise you on how to recover it.