METIS M311 / M322 / H311 / H322 (17-pin)
SensorTools Software
26
An identified pyrometer directly shows the current
measuring temperature in the preselected measuring
mode (2C for 2-color temperature, C1 or C2 for 1-color
temperature (channel 1 or channel 2) in this connec-
tion window as well as in the control window, and
graphically in the main window in the graph-tab as well
as the internal device temperature (DT).
Essential measuring parameters
can be entered via
basic device-specific input fields:
The
emissivity slope
must be taken into account for
measurements in the 2-color mode
(2C) and
adapted for the respective measuring material in order to obtain correct measured values (also see
The
emissivity
must be taken into account for
measurements in 1 channel mode
(C1 or C2) during
the measurement, and adapted at the pyrometer for the respective measuring material in order to ob-
tain correct measured values (also see
The
setpoint
for the PID controller, see
-
Via the
PID start button
the control process can be started directly.
-
The control parameters can be defined in the
controller window
Further measurement parameters
can be found here:
Opens a separate
control window
) to access
all
measurement-
relevant functions and settings when switching to other tabs (e.g. the graph view).
Device settings
are made here:
Via the
device configuration
basic device settings are made (see
).
The following settings can also be accessed from the "Device configuration", but can be found here
again for quick access:
The
data collection settings
control the transmission of measured data from the py-
Service functions
are used to back up, print, or restore device parameters, or create
service files for remote diagnostics in the event of a problem (see
Interface
) allows the display and / or setting of
-
the set
interface RS232
or
RS485
(changeover directly on the device)
-
the
baud rate
-
the
device address
(see
-
the
interface
or
answer delay
The recording function is used, as also in the graph window, to record the transferred meas-
ured value data for subsequent analysis to the hard disk (also see under
).
opens the
The
playback button
starts the
SensorTools
) and represents the last temporarily
stored measurement data, since the device was connected. These are not stored on hard disk as in
recording, only buffered. The amount of data depends on the value "Maximum records in memory"
(see
). Each measurement storage consists of a data set: measurement
2C: Under
DT: 31°C
2C: Measuring tempera-
ture
(at temperatures
below beginning of the
temperature range, the
software shows
“Under”)
DT: Device temperature
Measuring temperature
Identified device
Device configuration,
data collection settings,
service functions and
interface settings
Status of the digital in-
puts / outputs (see
Input fields
Recording / playback