METIS M311 / M322 / H311 / H322 (17-pin)
Technical Data
48
9.3
Composition of the Type Number
The equipment version of the pyrometer is encoded in the 22-digit model number as follows
(Example type number M311 with 600
–1300°C; laser targeting light, motorized focus, optics…):
Digit
01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22
Expl.:
M
3
1
1
0
6
0
0
1
3
0
0
1
5
8
1
3
0
4
2
3
A
Digit
Indication of the various points
Code
Meaning / Example
01-02
Series M3
M3
Metis M3
Series H3
H3
Metis H3
03-04
Spectral range
11
0.75
–1.1 µm (2 x Si)
0.78 + 0.99 µm (2 x Si)
22
1.45
–1.8 µm (2 x InGaS) or
1.4+1.64 µm (2 x InGaS)
05-08
Beginning of temperature range [°C]
0600
600°C
09-12
End of temperature range [°C]
1300
1300°C
13
Sighting
1
Laser targeting light
2
View finder
4
TV color camera
14
Serial interface
2
RS485
5
RS232 / RS485 (switchable)
15
Lens type
1
Fixed focus optics
2
Focusable optics
3
Fiber optic cable 0.2 mm Ø
with focusable optics
4
Fiber optic cable 0.4 mm Ø
with focusable optics
8
Motorized focus
B
Heavy duty stainless steel sensor head
for 0.2 mm fiber with OQ25
C
Heavy duty stainless steel sensor head
for 0.4 mm fiber with OQ25
16-17
Minimum response time
13
<1 ms
81
<80 µs
18
Configuration
7
17 pin connector (PID controller, 4 digital
inputs, 2 digital outputs, 1 analog input
0
–10 V, no display, no push button)
19
Digital display
0
Without display
20
Analog outputs
2
2 analog outputs
21
Digital input / digital output
4
4 digital 1 analog input
+ 2 digital outputs
(with 17 pin connector)
22
Measuring distance / optics
A
Optics A (340-3000 mm, integrated optics)
B
Optics B (240-3000 mm, optical fiber)
C
Optics C (120-500 mm, optical fiber)
X
Special optics