METIS M311 / M322 / H311 / H322 (17-pin)
Electrical Connection
11
Setpoint input via external voltage 0...10 V: This voltage
can also be made via the 10 V reference voltage output
and a 2 kΩ potentiometer (connected as a voltage divider,
the internal resistance of the input is 40 kΩ).
2 k
Ω
M3 / H3
Internal resistance
40 k
Ω
0 V
+10 V reference voltage
Voltage
divider
Power supply
0-10 V
White-green
Brown-green
Brown
Analog input
(input 5)
:
About a 0-10 V supply it can be adjusted ex-
ternally:
-
The emissivity slope
(0 V = 0.8; 10 V = 1.2)
-
The emissivity for every channel
(0 V = 0.05; 10 V = 1.2)
-
The setpoint for PID controller
(0 V = zero scale temperature,
10 V = full scale temperature)
-
Measuring distance at M3 devices with mo-
torized focus (0 V = shortest distance, 10 V
= widest distance)
4.1.4 Serial Interface RS232 / RS485
(M3: switchable RS232 / RS485; H3: only RS485)
The serial interface is used for digital communication of the pyrometer with another computer, for exam-
ple a PC for data transmission to the software
SensorTools
. Via interface always all temperatures are
transferred (2-color and one channel temperatures) as well as device information and parameters.
The maximum transmission speed (in Baud) is limited by the cable length, it is halved with each dou-
bling of the transmission path.
RS232:
about 7 m cable length with 19.2 Bd. Adjustable are values from 4.8 to 115.2 kBd.
RS485:
about 2 km with 19.2 kBd. Adjustable are values from 4.8 to 921.6 kBd.
Connecting one pyrometer via
RS232 or RS485:
In a short RS232 or RS485 connection to the mas-
ter (computer receiving the data), the pyrometer is
connected directly as a point-to-point connection
with the master.
It is advantageous to connect all interface cables
in order to avoid reflections.
Connecting several pyrometers via RS485:
For a reflection-free operation with longer cables, pay attention to the correct cable termination. Termi-
nation at the physical bus is at the beginning and at the end.
Master at
the beginning:
Master in
the middle:
Master
A (-)
B (+)
DGND
12
0
Ω
... Pyrometer n
Terminating
resistor 120 Ω
Pyrometer 2
Pyrometer 1
Terminating
resistor 120 Ω
in the master
or manually
A
B
GND
A
B
GND
A
B
GND
A
B
GND
GND
B
A
B
A
12
0
Ω
... Pyrometer n
Terminating
resistor 120 Ω
Pyrometer 1
A
B
GND
B
A
B
A
12
0
Ω
Master
A (-)
B (+)
DGND
A
B
GND
Terminating
resistor 120 Ω
RxD
TxD
GND
Pin 2
Pin 3
Pin 5
A (-)
B (+)
GND
RS485
(Master /
PC)
RS232 (PC)
Pyrometer
Pyrometer
A (-)
Brown-grey
B (+)
White-yellow
B (+)
White-grey
A (-)
Brown-yellow
TxD
Brown-grey
RxD
White-yellow
RxD
White-grey
TxD
Brown-yellow
DGND
White-pink
DGND
Brown-pink
DGND
White-pink
DGND
Brown-pink