METIS M311 / M322 / H311 / H322 (17-pin)
Pyrometer Parameters / Settings
24
6.11 Interface Type
(only M3)
Provides the interface to RS232 or RS485 (see
Serial Interface RS232 / RS485
).
6.12 Baud Rate
Serial Interface RS232 / RS485).
RS232: 4.8 to 115.2 kBd
RS485: 4.8 to 921.6 kBd
6.13 Temperature Unit
The temperature display can optionally be displayed in °C or °F.
6.14 Address
In bus operation with RS485 a device address between 00 and 97 can be assigned (see
Serial In-
terface RS232 / RS485
).
6.15 Interface Delay / Answer Delay
During operation of the pyrometer via RS485, it may happen that the connection is not fast enough to
detect the response of pyrometer time to a command from the master, e.g. because the line is still busy
with sending before receiving. This sometimes occurs in older PCs or interface adapters or by slow
switching times of interface adapters and manifests itself in transmission errors of interface commands,
so errors when parameters are changed or of measured values transmission.
In this case a delay for delaying the data transfer can be entered, the value specifies the main circulat-
ing time and can be set from 00-20. The transmission of commands to the pyrometer is thus somewhat
delayed, but it can work with existing peripherals.