METIS M311 / M322 / H311 / H322 (17-pin)
SensorTools Software
30
7.2.2 Control Window
Measuring mode
2C
means that the
2-color temperature
is displayed.
In the measured temperature selec-
tion field it can be changed to
measuring temperature
of
chan-
nel 1 or channel 2
.
The setpoint of the
PID controller
is always displayed.
If a ramp time is defined, the set-
point can always display the updat-
ed value when the buffer mode with
status information is selected in the
data collection settings (see
The
internal
device temperature
(DT)
depends on the ambient temperature and the device self-heating.
-
M3 models are permissible up to maximum 80°C.
-
H3 models are permissible up to maximum 60°C.
-
The laser targeting light is deactivated at a device temperature from 60°C, the camera module of
M3 devices from 55°C to prevent its overheating). Below that temperature it can be used again.
The status of the
digital inputs / outputs
is indicated by an LED symbol:
-
Input (IN) or output (OUT) active
-
Input (IN) or output (OUT) inactive
-
Input (IN) or output (OUT) not activated (no function)
Adapt parameters:
Parameter changes are immediately transferred to the pyrometer.
- Device parameters
- The
response time t
90
(see
can be adapted to the measurement conditions.
Selecting "Minimum" the pyrometer works with its shortest possible response time of <1 ms (M3)
or <80 µs (H3). Under "User definded" times between 1 ms (M3) or 0.1 ms (H3) and 10 s can be
entered.
-
Storage mode
): switches on (and off) the
peak picker
and allows clear settings for the
activated peak picker.
- Measurement parameters:
The displayed measuring parameter depends on the measuring
mode. In case of ratio temperature measurements, the emissivity slope can be set (see
); in the
case of 1 channel measurements, the emissivity input field is available first.
-
The setting field can be expanded to display the measurement parameters for all channels.
-
To display further sub-parameters, each individual parameter can be extended
-
Emissivity slope
): Adjustment of the surface
properties in 2-color mode.
-
Switch-off level
): Input when the measurement
is turned off by low signal strength.
-
Switch-off verification
): Input when the
switch-off is actually made.
-
Emissivity
(see
): Adjustment of the surface proper-
ties in 1-color mode.
-
Transmittance
): Transmittance of a window
-
Spot size fill factor
): For 1-color measure-
ments, for targets smaller than the spot size.
Quick adjustment of parameters
: Changes the currently selected
parameter (e.g. click with the mouse in a field or select it us-
ing the Tab key in sequence of the fields).
Controller parame-
ters (optional)
Device configuration
Laser targeting light on-
off / camera settings
Status of the digital
inputs / outputs
Measuring mode
selection
Device temperature
Temp. measuring mode
Device parameters
Measurement pa-
rameters quick save
Parameter quick
adjustment
Measured temperature
Set point temp.
(devices with PID
controller)
Measurement parame-
ters (extendable)
Motor focus adjustment
(optional)
Show measuring
parameters of all
channels
Further settings
for measuring
parameters