Reading Parameters:
Function (Read)
Command
Output
Response time t
90
AAez<CR>
X
(1-digit. dec.; 0...6)
0= intrinsic time constant of the device
1=0,01 s
4=1,00 s
2=0,05 s
5=3,00 s
3=0,25 s
6=9,99 s
Error status
AAfs<CR>
XX
(2-digit. hex.)
Bit 0=1: Signal unit failure
Bit 1=1: Devicemp1.failure
Device address
AAga<CR>
XX
(2-digit. dec.; 00...99)
Device temperature
AAgt<CR>
XX
(2-digit. dec.; 00...98 °C)
Maximum storage
AAlz<CR>
X
(1-digit. dec.) 0 = Maxi off
1=0,01 s
5=5,00 s
2=0,05 s
6=25,00 s
3=0,25 s
7=external clear contact
4=1,00 s
8=automatic clear
Basic range
AAmb<CR>
XXXXYYYY
(8-digit. hex.)
XXXX=lower range limit in °C
YYYY=upper range limit in °C
Sub range
AAme<CR>
XXXXYYYY
(8-digit. hex.)
XXXX=lower range limit in °C
YYYY=upper range limit in °C
Parameter general
AApa<CR>
(11-digit. dec.)
1.-2.digits (20...99 or 00)
3. digit (0...6) t
90
Response time
4. digit (0...8) t
cl
Maximum storage
5. digit (0...1) Analog output
6.-7. digit (00...98) Device temperature
8.-9. digit (00...99) Device address
10. digit (0...4 or 6) Baud rate
11. digit (0) always 0
Serial no.
AAsn<CR>
XXXXX
(5-digit. dec.)
Max. Device temperature
AAtm<CR>
XX
(2-stell. dec.; 00...98 °C)
Device type and
software version
AAve<CR>
XXYYZZ
(6-digit. dec.)
XX = 11 (
Metis MS/MI-WW
)
YY = Month of software version
ZZ = Year of software version