Metis MS /MI WALZWERK

 

 

 

Device internal temperature (command = gt) 

 

The  device  internal  temperature  can  be  read  via  the  interface.    The  internal 

temperature  lies  some  degrees  above  the  environment  temperature  in 

continuous operation because of the intrinsic heating due to the electronics. 

Bus address (command = ga) 

This parameter is important for bus operation with more than one RS485 device. 

In this case each device gets its own address. At RS232 devices ADR is usually 

set to 00. 

Every  device  communicates  via  the 

global  address  99

,  (independently  of  its 

own  address).  This  is  the  way  to  get  communication  with  a  device  where  the 

address is unknown. 

For RS485 operating mode, you may change parameters of all bus pyrometers 

using the 

group  address  98.

 The devices will not respond at this address. So 

only setting commands like em, lz, ez and as are available. 

Baud rate (command=br) 

 

Here  you  pre-set  the  speed  of  the  serial  interface  (in  kBd)  depending  on  your 

hardware. Following values are possible: 

 

1200 Bd, 2400 Bd, 4800 Bd, 9600 Bd, 19200 Bd and 57600 Bd and 115000Bd. 
Default setting is: 19200 Bd 

 

9. Operation via serial interface 

With operation via serial interface there are several features: 

- reading the measured object temperature 

- reading the limits of the basic range 

- diagnostics (signal unit OK?, device internal temperature measurement OK?) 

- programming of a shorter sub-range within the basic range (with software 

SensorWin.exe) 

For  interface  operation  you  may  either  use  the  program  "SensorWin.exe"  (see 

chapter  10)  with  a  terminal  program  (Term.exe)  or  you  write  your  own 

RS232/RS485 communication software for data exchange with the device (see 

Appendix referring data format and protocol). 

 

Summary of Contents for Metis MI 16

Page 1: ...Beschreibung und Bedienungsanleitung Technical Documentation and Operation Manual Pyrometer Metis MS09 MI16 MI18 Walzwerkausf hrung Hot Rolling Mill Equipment...

Page 2: ...stellung der Parameter Setting of Parameter 13 37 7 Miniaturschalter und Men f hrung Miniature switches and menue driven operation 13 37 8 Ger te Parameter Device Parameter 18 42 9 Bedienung ber Schni...

Page 3: ...uszuf hren und dabei m gliche Gefahren erkennen und vermeiden k nnen Definition f r Fachkr fte laut IEC 364 Sicherheitshinweise Lesen Sie diese Bedienungsanleitung sorgf ltig und gewissenhaft um Siche...

Page 4: ...ber die Schnittstelle k nnen sowohl Messwerte ausgelesen als auch Ger teparameter wie z B der Emissionsgrad fernverstellt werden Das Ger t nutzt alle Vorteile der digitalen Messwertverarbeitung und h...

Page 5: ...ette mit PC Auswertesoftware und Imbusschl ssel 2 5mm ausgeliefert Ein Anschlusskabel ist nicht im Lieferumfang enthalten und muss extra bestellt werden s Zubeh r Bestellnummernschl ssel Metis MS MI W...

Page 6: ...sionsgrad Erfassungszeit 0 4 20 mA Umschaltung f r Analogausgang Ger teinnentemperatur Adresse Baudrate Teilmessbereichsgrenzen innerhalb des Grundmessbereichs Aufl sung 0 1 C an der Schnittstelle 0 1...

Page 7: ...C MB 13 315 0 8 16 720 1 6 16 350 1800 C MB 18 315 0 8 7 720 1 6 7 Metis MI 18 Messabstand a mm Messfeldgr e M mm Apertur D mm 120 550 C MB 5 315 1 6 16 720 3 3 16 160 800 C MB 8 315 0 8 16 720 1 6 1...

Page 8: ...ei w hlbar bei Bestellung angeben W L nge Infrarot Messkopf Messwertaufnehmer 249mm R L nge Sp lrohr Auswahl R 225mm bei Optik a 315 R 630mm bei Optik a 720mm bei Bestellung angeben a Messabstand Abst...

Page 9: ...9 Metis MS MI WALZWERK Metis Metis Metis Metis MS MI WW Versorgungselektronik Metis Metis Metis Metis MS MI WW Versorgungselektronik Profibus...

Page 10: ...uftsp larmatur und an der anderen Seite einen M16 8 pol Stecker Anschluss Versorgungselektronik An der R ckseite ist der 12 polige Anschlussstecker Ansicht auf die Stiftkontakte Stecker Pin Bedeutung...

Page 11: ...n 4 B gelb 5 H grau 6 J rosa 7 M orange 8 M orange 9 G rot 10 F schwarz 11 C violet 12 D grau rosa 13 E rot blau Pin Farbe Bedeutung K wei 24 V Versorgungsspannung oder 24 V AC A braun 0 V Versorgungs...

Page 12: ...asse f r Schnittstelle G Der 9 polige SUB D Stecker am PC hat folgende RS232 Belegung Pin Bedeutung 2 Receive RxD Empfangs Daten vom Pyrometer zum PC 3 Transmit TxD Sende Daten vom PC zum Pyrometer 5...

Page 13: ...f hrung Um Ger teeinstellungen vorzunehmen m ssen die vier Innensechskantschrau ben der R ckwandplatte gel st werden Benutzen Sie hierf r einen Schl ssel mit SW 2 5 mm Achtung Wird das Ger t bei anlie...

Page 14: ...ebenfalls nur bei Profibusger ten ben tigt und kann bei Standardger ten ohne Profibus beliebig eingestellt sein Men f hrung optional Das Ger t verf gt zus tzlich ber 4 Bedientasten Von links nach rech...

Page 15: ...mt nur zum Einsatz um in ein Untermen zu springen z B bei Reglern etc Man erkennt ein Untermen daran dass der Name mit dem Kleiner und Gr er Zeichen umrahmt ist Zum Beispiel REGLER Stromausgang pr fen...

Page 16: ...rameter Maximalwertspeicher Display MAXI OFF Parameter Ausgang Display AUS 0 20mA Parameter Messbereichsanfang Display MBA 300 C Parameter Messbereichsende Display MBE 1300 C Parameter Adresse Display...

Page 17: ...n Die Dichtung muss sich leicht zwischen Geh use und Deckel herausdr cken F r Ger tesch den durch eingetretene Fremdstoffe oder vorschriftswidrige Befestigung des Geh usedeckels kann keine Gew hrleist...

Page 18: ...stark verzundert 0 93 0 85 0 9 Eisen Walzhaut 0 88 0 8 0 88 Eisen fl ssig 0 3 0 2 0 25 Schlacke 0 85 0 8 0 85 Aluminium 0 15 0 1 Chrom blank 0 28 0 32 0 25 0 3 Messing oxidiert angelaufen 0 65 0 75 0...

Page 19: ...messen Wird eine von 0 abweichende L schzeit in Sekunden eingegeben wird der Maximalwert ermittelt und in einem Doppelspeicher festgehalten Nach der gew hlten L schzeit wird jeweils ein Speicher gel s...

Page 20: ...telle kann auf folgende Werte festgelegt werden 1200 Baud 2400 Baud 4800 Baud 9600 Baud 19200 Baud 57600 Baud und 115000 Baud Standardm ig ist 19200 Baud eingestellt 9 Bedienung ber Schnittstelle Bei...

Page 21: ...Kommunikation mit dem Ger t Ger t ber Anschlusskabel mit 24V versorgen und ber 9 poligen Stecker falls erforderlich ber 9 auf 25 Adapterst ck mit einer seriellen Schnittstelle COM1 bis COM9 des PC s...

Page 22: ...XXX 5 stell dez in C letzte Stelle ist Nachkommastelle Sonderfall 88880 Temp Overflow Messwert mehrfach AAmsXXX CR XXX Anzahl der Messwerte Es werden entsprechend XXX Messwerte gesendet Siehe oben Par...

Page 23: ...l schen 4 1 00 s 8 automat Grundmessbereich AAmb CR XXXXYYYY 8 stell hex XXXX Messbereichsanfang in C YYYY Messbereichsende in C Teilmessbereich AAme CR XXXXYYYY 8 stell hex XXXX Messbereichsanfang in...

Page 24: ...01 s 4 1 00 s 2 0 05 s 5 3 00 s 3 0 25 s 6 9 99 s ok Ger teadresse AAga CR XX 2 stell dez 00 99 00 97 einstellbare Ger teadressen 99 Globale Adresse mit Antwort 98 Globale Adresse ohne Antwort Globale...

Page 25: ...d thereby recognise and are able to avoid any possible dangers definition of qualified employees according to IEC 364 Safety Notice Read the operating instructions carefully and thoroughly in order to...

Page 26: ...le RS485 The Interface enables the user to read out measured data as well as to set hardware parameters e g the emissivity remotely The device uses all advantages of digital measurement and has compar...

Page 27: ...MS MI MS MI MS MI WW WW WW WW Factory settings Analog output 0 20 mA Device baud rate 19 2 kBd Emissivity 100 Response time t90 1ms Device address 00 Erase maximum storage Storage off Sub range limit...

Page 28: ...1 12 13 14 15 16 17 18 19 20 21 22 M I 1 8 0 5 5 0 1 4 0 0 0 5 1 1 3 3 2 0 0 Serial M Metis Metis Metis Metis Detector S Silizium I InGaAs Center of wavelength 09 0 9 m 16 1 6 m 18 1 8 m Lower limit r...

Page 29: ...sured quantity emissivity recording time 0 4 20 mA select for analog output device temperature address baud rate measuring sub ranges within the limits of basic range Resolution 0 1 C at the interface...

Page 30: ...C MB 10 315 1 6 16 720 3 3 16 300 1300 C MB 13 315 0 8 16 720 1 6 16 350 1800 C MB 18 315 0 8 7 720 1 6 7 Metis MI 18 Meas distance a mm Spot size diameter M mm Aperture D mm 120 550 C MB 5 315 1 6 1...

Page 31: ...Selection 5m 10m freely selectable with order W Length Infrared sensor head readings recorder 249mm R Length air purge Selection R 225mm at a 315 R 630mm at a 720mm with order a Measuring Distance Dis...

Page 32: ...32 Metis MS MI WALZWERK Metis Metis Metis Metis MS MI WW Supply unit electronics Metis Metis Metis Metis MS MI WW Supply unit electronics profibus...

Page 33: ...ing on one end and an M16 8 pin socket connection supply unit electronics on the other end Pin Layout On the rear panel is a 12 pin connector Pin Layout Connector Pin Function K 24 V supply voltage or...

Page 34: ...ack 11 C Violet 12 D grey pink 13 E red blue Pin Color Function K white 24 V supply voltage or 24 V AC A brown 0 V supply voltage L green Iout analog current output B yellow Iout analog current output...

Page 35: ...D Transmission data from PC to Pyrometer 5 Signal Ground GND Zero Benchmark Connector Cable One End has a 12 Pole M23 Cable Socket and the other End an 8 Pole M16 Cable Plug AK86 05 Standard Length 5m...

Page 36: ...ll as B1 and B2 are connected inside the 12 pin connector of the connecting cable order no AK10 XX to avoid any long transmission lines and thereby avoid reflections onto the bus Pyrometer 1 Pyrometer...

Page 37: ...se a 2 5 mm hex Allen wrench When you have loosened the four screws it is possible to pull the rear panel like a drawer Inside you will find 10 miniature switches and a button Warning Use caution when...

Page 38: ...ameter will be chosen again Now select the desired parameter to change it just push the UP or DOWN button for the desired menu parameter The parameter value is now flashing If the desired value is sho...

Page 39: ...ll set the output back to the standard measuring mode Otherwise the pyrometer itself will automatically return to the measuring mode after approx 45 seconds Configuration with SensorWin The program Se...

Page 40: ...Parameter Maximum storage Display MAXI OFF Parameter Output Display OUT 0 20mA Parameter Zero Scale Display ZSC 300 C Parameter Full Scale Display FSC 1300 C Parameter Address Display ADD 00 Parameter...

Page 41: ...manufacturer is not responsible for device malfunctions caused by dirt or corrosions inside the device case Please make sure that the cover is always mounted appropriately otherwise your warranty may...

Page 42: ...93 0 85 0 9 Iron rolling skin 0 88 0 8 0 88 Iron molten 0 3 0 2 0 25 Slag 0 85 0 8 0 85 Aluminum 0 15 0 1 Chromium bright 0 28 0 32 0 25 0 3 Brass oxidized tarnished 0 65 0 75 0 65 0 7 Bronze 0 03 0...

Page 43: ...nt measurement errors Longer response times may be useful for averaging over fast changes of the object temperature Clear time clear mode for maximum storage command ez At clear time 0 the maximum val...

Page 44: ...ress 98 The devices will not respond at this address So only setting commands like em lz ez and as are available Baud rate command br Here you pre set the speed of the serial interface in kBd dependin...

Page 45: ...g communication with the pyrometer Apply 24 V supply to the device with the connecting cable and connect the 9pin connector to a serial PC interface COM1 or COM2 After the start of the SensorWin softw...

Page 46: ...R XXXXX 5 digitl dec in C Last digit is positioned after the decimal point Special case 88880 Temp Overflow Temperature multi AAmsXXX CR XXX Number of values XXX measured values will be sent See ms co...

Page 47: ...t 4 1 00 s 8 automatic clear Basic range AAmb CR XXXXYYYY 8 digit hex XXXX lower range limit in C YYYY upper range limit in C Sub range AAme CR XXXXYYYY 8 digit hex XXXX lower range limit in C YYYY up...

Page 48: ...the device 1 0 01 s 4 1 00 s 2 0 05 s 5 3 00 s 3 0 25 s 6 9 99 s ok Device address AAga CR XX 2 digit dec 00 99 00 97 regular device address 99 global address with response 98 global address without r...

Page 49: ...49 Metis MS MI WALZWERK...

Page 50: ...50 Metis MS MI WALZWERK...

Page 51: ...51 Metis MS MI WALZWERK...

Page 52: ...herm GmbH Sensortherm GmbH Sensortherm GmbH Firmensitz Headquarter Hauptstra e 123 D 65843 Sulzbach Ts Tel 49 0 6196 64065 80 Fax 49 0 6196 64065 89 Email info sensortherm de Internet www sensortherm...

Reviews: