Setting Up ReDS System
44
Tap Next to continue.
Adjust Back Center Strip
Figure 30: Adjust the Back Center Strip
1
While the patient is holding the front buckle steady on the sternum, carefully center the
back of the sensor vest by aligning the lower end of the Back Center Strip with the
patient’s spine. To verify, use your fingers to feel the spinal column in the lower part of the
vest near the back sensor. Once centered, hold the back piece in place.
Figure 30: Centering the Back of the Sensor Vest
Align the lower end of the
Back Center Strip with the
patient's spine.
Use your fingers to feel that
the spine is just below this
area of the Back Center Strip.