The “Read” button (upper-left side first icon) allows a user to retrieve the stored data.
The “Write” button (upper-left side second icon) has one of the most important role in this
software interface. Because every single and each configured or customized settings will
be saved by clicking “Write” button. When a user configures the instrument`s settings,
“Write” button will be clicked and message will pop-up. Click “Yes”.
Calibration is the comparison of measurement values delivered by a device under test
with those of a calibration standard of known accuracy. To perform the calibration using
software, follow the below:
All recent 30 logs will be stored in the device and
will be automatically erases one by one from the
earliest logs when new event occurs. There are
two types of event logs, “Event Log” and “Event +
Data log” are available to download. Choose the
log and click “Download” button. The log files will
5.1.2. Write
5.1.3. Calibration
5.1.4. Log
1. Connect the unit to the PC using instrument`s
USB
2. Plug the calibration cap (not for Fresh air
calibration) and open the software
3.
Click “Calibration” (middle-left side icon) and
wizard will come up
4.
Choose the calibration gas type and click “Start”
5. The time for Fresh Calibration is 10seconds while
for N
₂
and CO
₂
is 90 seconds
be downloaded and created by unit`s Serial number and will be in “.csv” format. However,
clicking “Erase” button will clean all the logs from the storage of device and cannot be
recovered.
To upgrade the latest firmware version of iGas
Detector CO
₂
,
follow the below:
1.
Click “Browse” button and navigate to the
firmware location
2.
Choose the firmware and click the “Open” button
3.
Click “Write” to begin upgrading process
4. When upgrade is finished, power off the device
and turn it on
5.
The “F-UP” → “boot” message will come up and
upgrade is complete
5.1.5. Upgrade (Firmware)
14
Note:
-
Pressing “Cancel” button during the
upgrading process will cancel and
close the Firmware Upgrade Wizard.
5.1.1. Read
5. Software Manager