background image

Gas Type

Fresh Air (O

)

N

CO

Concentration

20.9%vol

99.99%vol

20,000ppm, 2%vol

Calibration Gas

1. In the measurement display, press and hold down the pushbutton and power button to 

enter the setting mode.

2. Press the pushbutton until the CAL mode is displayed. 
3. In the CAL mode, press the power button for 2 seconds to enter the mode
4.

In the arm’s length like the right image, hold the detector and for 2 seconds and press 
the power button to perform the fresh air calibration. 

5. Once the calibration is successful, the success message (

V

) mark is displayed. But, if 

it fails, the FA  message(

X

) mark is displayed. 

6. After the successful calibration, the baseline is set to 

400ppm(0.04%vol). 

Fresh Air calibration

1

CAUTION

-

Initial calibration is performed on all devices prior to shipment. 

-

The detector has fresh air calibration, N2 calibration and CO2 span calibration.

The sensor takes the last calibration, so if you have CO2 gas, we only recommend CO2 calibration.
If you don't have CO2 gas, N2 or fresh air calibration are also available for calibration.

- Check with your safety managers to ensure calibration accessibility. 

-

Note that Fresh air calibration should be performed out of your office in a fresh air and in the arm’s 
length to avoid the CO2 interference from breathing. 

-

Before calibration, move to a fresh air, which is free of toxic and combustible gases. 

-

If calibration fails, perform re-calibration again. If the repeated calibration continues to fail, contact 
authorized safety managers or distributors. 

-

The auto fresh air calibration can be set by the software manager. 

Once activated, the fresh air calibration is activated every 3 days. (Refer to page.13)

4.2. Calibration 

4.  Setting Mode

CAUTION

-

The presence of CO

in uncontaminated ambient air is around 

400ppm (0.04%) and it exceeds over 1000ppm (0.1%) in any 
enclosed environments such as offices, workplaces, room, etc. 
as a result of respiration, heat from electronic devices and poorly 
ventilation. Therefore, to do fresh air calibration, the calibration 
must be in outdoor and well away from a building and the 
breathing zone. Ensure to hold the detector in the arms’ length 
like the right image. 

10

Summary of Contents for iGas Detector CO2

Page 1: ...Portable Single Gas Detector User s Manual...

Page 2: ...sensor regularly by the gas concentration exceeding alarm set points Test LED audio and vibration manually If the temperature changes sharply during use of the device e g indoors vs outdoors the valu...

Page 3: ...etting Mode 4 1 Alarm Adjust alarm setpoints 4 2 Calibration 4 3 Clr max 5 Software Manager 6 Maintenance 4 5 6 8 9 10 11 13 16 7 Specification 18 8 Battery Charging 18 3 3 Menu Tree 7 4 4 Clr STEL an...

Page 4: ...ration Remaining Battery 1st Alarm Display 2nd Alarm Display Alarm Condition Max Value Short Term Exposure Limit Time Weighted Average Test Success Test Fail Over Limit End of Test Part Description LC...

Page 5: ...r 1 Initial setting error Err 2 Sensor error Err 3 Memory error L bat Low battery Err marks OFF 2 1 Bump test 1 Before daily use users are required to perform bump test to see a sensor responds to a C...

Page 6: ...oint STEL alarm setpoint TWA alarm setpoint Firmware version Calibration concentration To move to the next menu press the pushbutton At the last step press pushbutton or do not push any button for 10...

Page 7: ...Factory Default Test 2 sec Fail Click Click Measuring mode Click Next mode 2 sec Setting process Cal Setting Process Value Clear Process Value Clear Process Setting process Setup Testing progress 2 se...

Page 8: ...ncentration Unit vol ppm Concentration unit conversion Init Reset Test Self Bts Self test Bump test 8 In the setting mode users can adjust setpoints perform calibration and reset previous values 1 To...

Page 9: ...bration per second Adjust alarm setpoints To enter the setting mode press and hold the pushbutton and power button simultaneously for two seconds In the alarm setting icon press and hold down the powe...

Page 10: ...e CO2 gas N2 or fresh air calibration are also available for calibration Check with your safety managers to ensure calibration accessibility Note that Fresh air calibration should be performed out of...

Page 11: ...peak concentration in the detector follow below steps 1 Press the pushbutton power button simultaneously and the pushbutton until Clr max is displayed 2 Press the power button to clear the peak value...

Page 12: ...Press the pushbutton until Test is displayed 2 Press the power button for three seconds In the SELF display press the power button for three seconds to activate the self test While it s activated the...

Page 13: ...e message Unit Adjust measuring unit by PPM or vol Stealth Disable the alarm buzzer and LED for a special occasion Calibration Use Disable Enable the calibration access for a special occasion Auto Cal...

Page 14: ...e log files will 5 1 2 Write 5 1 3 Calibration 5 1 4 Log 1 Connect the unit to the PC using instrument s USB 2 Plug the calibration cap not for Fresh air calibration and open the software 3 Click Cali...

Page 15: ...e unit Test order LED Beep Motor Flash Sensor End Factory Default F Reset original settings and specifications Time Write T To set a time by user location see 8 2 3 1 Power OFF P Turn off the device I...

Page 16: ...safety function Replacing the sensor and battery should be performed by authorized sellers agents distributors or safety managers The sensors published by SENKO should be used for replacement Product...

Page 17: ...bration with CO2 2 vol Before use stabilize the detector for 5 minutes 6 1 Sensor Replacement 6 2 Battery Replacement 1 Move to a fresh air environment and power off the detector 2 Remove the back cas...

Page 18: ...0g Operating Life 14 days Expected in measuring mode without alarm Event Log Recent 30 alarms Approval EMC directive 2014 30 EU Gas Measuring range Low Alarm High Alarm Resolution CO2 0 5 0 vol 0 5000...

Page 19: ...nufacturer is not liable under this warranty if its testing and examination disclose that the alleged defect in the product does not exist or was caused by the purchaser s or any third party s misuse...

Reviews: