CAUTION
-
Do not change the calibration concentration unless dealers or safety managers authorized by
SENKO give the permission to change to another calibration concentration.
-
Use the regulator with a flow rate of 0.5LPM(Liters per minute) of a gas cylinder.
1. Press the pushbutton to move to span calibration.
2. Plug the calibration cap to the detector and connect the calibration cylinder with
CO2 (20,000ppm)
3. Press the power button and release CO2 gas.
4. After 90 seconds, when the calibration is successful, success message(
V
) appears.
If N2 calibration fails, fail message(
X
) appears.
To delete the measured peak concentration in the detector, follow below steps.
1. Press the pushbutton & power button simultaneously and the pushbutton until Clr(max)
is displayed.
2. Press the power button to clear the peak value.
3. After the successful activation, SUC (
V
) mark is displayed. If it fails, FA (
X
) mark is
displayed.
4.3.
Clear
max
N2 Calibration
1.
In the “CAL” mode, press the pushbutton to move to
the “N2” calibration.
2. Plug the calibration cap to the detector and connect the
calibration cylinder with N2 (99.9%vol)
3. Press the power button and release the N2 gas.
4. After 90 seconds, when N2 calibration is successful,
success message(
V
) appears. But, If N2 calibration is
failed, FAIL message(
X
) appears.
2
Span Calibration
3
11
CAUTION
-
Use the regulator with a flow rate of 0.5LPM(Liters per minute) of a gas cylinder.
4. Setting Mode